DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
83 of 180
12914 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12915 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12930 - SLOW SCAN: Unexpected Retry during Current Test
12931 - SLOW SCAN: Unexpected Retry during Velocity Test
Description: The command MOVE TO EXTRACTION BAR REFERENCE SENSOR
did not occur. The command is executed again.
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Description: The command MOVE TO THE PLATE GUIDE POSITION did not occur.
The command is executed again.
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Description: The system was not able to move to the position selected in the
diagnostics. The command is executed again.
Cause:
The SLOW SCAN CONTROLLER BOARD malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Description: The system was not able to move to the position selected in the
diagnostics. The command is executed again
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the