DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
88 of 180
14006 - DUPLEX CAM: Initialization Error due to the PLATE PRESENT
SENSOR S5 reading blocked when the SLOW SCAN AY is in the home
position
14008 - SLOW SCAN: Initialization Error
14009 - DUPLEX CAM: Initialization Error
14010 - Slow Scan Async Reset Error
Cause 1: The PLATE PRESENT SENSOR S5 operates correctly, but another malfunction
has occurred. This error occurs when initializing, either when the CR 825/850
SYSTEM is energized or when a mechanical subsystem initializes.
Action:
Check:
•
PLATE was removed from the CASSETTE
•
SLOW SCAN MOTOR operates correctly
•
obstruction in the path of the EXTRACTION BAR
Cause 2: The PLATE PRESENT SENSOR S5 malfunctioned.
Action:
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Cause:
The DUPLEX CAM malfunctioned.
Action:
Do the checkout for the
Cause:
The CR 825/850 SYSTEM sent a message to reset the SLOW SCAN AY.
Action:
None.