DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
82 of 180
12907 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12908 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12911 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
12913 - SLOW SCAN CONTROLLER BOARD A6: Unexpected Retry
Description: The FINE NUDGE DOWN command did not occur. The command is
executed again.
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Description: The PLATE did not move to the CASSETTE. The command is executed
again.
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Description: The SOFT RESET command did not occur. The command is executed
again.
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Description: The EXTRACTION BAR did not move to the mechanical adjustment
position one. The command is executed again.
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the