DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
92 of 180
15013 - Image Quality Warning
15014 - LASER power not calibrated
15015 - PLATE not calibrated
Description: The image quality might be wrong. The MASTER CENTRAL PROCESSING
UNIT BOARD A2 received a system error from the MOTION SYSTEM
CONTROL BOARD A1 and was automatically repaired. If a CASSETTE was in
the CR 825/850 SYSTEM when the error occurred, the TOUCH SCREEN
displays a message about a problem with the image.
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Do the checkout for the
MASTER CENTRAL PROCESSING UNIT BOARD A2
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Do the checkout for the
MOTION SYSTEM CONTROL BOARD A1
Cause 3: A light leakage occurred.
Action:
Cause 4: The INTERLOCK SWITCH S8 malfunctioned. The DOOR might be open during
scan.
Action:
.
Cause:
The LASER DIODE malfunctioned.
Action:
Do the calibration for the LASER DRIVER BOARD A17. See
.
Cause:
The CR 825/850 SYSTEM does not have the calibration data for all sizes of
PLATES.
Action: