DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
77 of 180
12125 - SLOW SCAN COIL BOARD A7 malfunctioned
12126 - Caution! SLOW SCAN MOTOR is overheated
12130 - SLOW SCAN CONTROLLER BOARD A6: Software Error
12131 - SLOW SCAN CONTROLLER BOARD A6 did not communicate with
MSC
Description: The SLOW SCAN AY stopped or did not move at all.
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Description: The SLOW SCAN MOTOR was using excessive power and is too hot.
Cause:
Malfunction in one or more of the following:
•
COIL BOARD A7
•
EXTRACTION BAR SENSOR S9
•
SLOW SCAN AY
Action:
1.
Check for an obstruction.
2.
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Cause 1: The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Cause 2: The MOTION SYSTEM CONTROL BOARD A1 malfunctioned.
Action:
Do the checkout for the