DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
98 of 180
16013 - Imaging System contains out-of-range noise
16014 - Imaging System did not zero or PMT did not calibrate
Cause 1: The MASTER CENTRAL PROCESSING UNIT BOARD A2 malfunctioned.
Action:
Do the checkout for the
MASTER CENTRAL PROCESSING UNIT BOARD A2
Cause 2: The PMT/DAS BOARD A5 malfunctioned.
Action:
Do the checkout for the
Description: The “peak-to-peak” signal was high when the CR 825/850 SYSTEM was
set to “AUTO-ZERO.”
Cause 1: The PMT/DAS BOARD A5 malfunctioned.
Action:
Do the checkout for the
Cause 2: Noise in the POWER SUPPLY PS1.
Action:
Do the checkout for the
Cause 1: The DOSIMETER malfunctioned.
Action:
Check that the DOSIMETER operates correctly.
Cause 2: The exposure on the PLATE is not correct during the calibration.
Action:
Do the calibration with another PLATE.
Cause 3: The X-RAY TUBE malfunctioned.
Action:
Expose the PLATE using a new X-RAY TUBE.