DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
78 of 180
12132 - SLOW SCAN CONTROLLER BOARD A6 did not communicate with
MSC
12201 - SLOW SCAN CONTROLLER BOARD A6: Programming Error
12202 to 12260 - SLOW SCAN CONTROLLER BOARD A6: “Memory
Command” Error
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Cause:
The SLOW SCAN CONTROLLER BOARD A6 malfunctioned.
Action:
Do the checkout for the
SLOW SCAN CONTROLLER BOARD A6 and COIL
Code
Description
12202 Error occurred during downloading to the SLOW SCAN CONTROLLER BOARD A6.
12203 Memory error on the SLOW SCAN CONTROLLER BOARD A6.
12204 Memory error on the SLOW SCAN CONTROLLER BOARD A6.
12216 Communication error between the SLOW SCAN CONTROLLER BOARD A6 and the
MOTION SYSTEM CONTROL BOARD A1.
12217 Communication error between the SLOW SCAN CONTROLLER BOARD A6 and the
MOTION SYSTEM CONTROL BOARD A1. Error “EM17 - Receive Buffer Overflow.”
12218 Error during downloading of code parameters for the SLOW SCAN CONTROLLER
BOARD A6.
12219 Error during downloading of code or parameters for the SLOW SCAN
CONTROLLER BOARD A6.
12220 Error “EM20 - New Bootstrap Program, Download all Boot and Program Modules.”
12221 Malfunction of the SLOW SCAN CONTROLLER BOARD A6.