DIAGNOSTICS
Error Codes
10DEC05
DG4825-1
Page
99 of 180
16016 - Malfunctioning PMT
16017 - Possible Light Leak
16018 - Possible Light Leak
16019 - Calibration could not find offset, rotate GALVO counterclockwise
16020 - Calibration could not find offset, rotate GALVO clockwise
Description: A PMT does not match the requirements of the diagnostic test.
Cause:
The PMT/DAS BOARD A5 malfunctioned.
Action:
Do the checkout for the
Cause:
The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
Action:
1.
Do the
to locate the source of the light leakage.
2.
Repair the light leakage.
Description: An error occurred during “AUTO-ZERO.” The PMTs detected a large
quantity of light.
Cause:
The DOOR is open, a PANEL is loose, or a light leakage occurred in the
CASSETTE HANDLING SYSTEM.
Action:
1.
Do the
to locate the source of the light leakage.
2.
Repair the light leakage.
Cause:
The calibration could not locate the “Offset.”
Action:
Rotate the GALVO counterclockwise.
Cause:
The calibration could not locate the “Offset.”
Action:
Rotate the GALVO clockwise.