Stratos Multi E401N
138
Error no. Type Message text
Notes/solution
F212
Time/Date
The time and date still need to be set.
Parameter Setting
System Control
Time/Date
F229
Info
Wrong Passcode
An incorrect passcode was entered.
F236
HART not Available,
Current too Low
HART is not available with currents smaller than 4 mA: Set current output I1
to 4-20 mA.
Parameter Setting
Inputs and Outputs
Current Outputs
Current Output I1
Output
Current Output/Relay Contacts
Error no. Type Message text
Notes/solution
B070
Current I1 Span
Current output 1: The selected span is too small/too big:
Parameter Setting
Inputs and Outputs
Current Outputs
Current Output I1
Check start/end.
B071
Current I1 < 0/4 mA
Current output 1: The current is below the permissible limit.
B072
Current I1 > 20 mA
Current output 1: The current is above the permissible limit.
B073
Current I1 Load Error
Current output 1: The current loop has been interrupted (open circuit) or
the load is too high: Deactivate or short-circuit unused current outputs.
B074
Current I1 Parameter
Current output 1: Check the configuration.
B075
Current I2 Span
Current output 2: The selected span is too small/too big:
Parameter Setting
Inputs and Outputs
Current Outputs
Current Output I2
Check start/end.
B076
Current I2 < 0/4 mA
Current output 2: The current is below the permissible limit.
B077
Current I2 > 20 mA
Current output 2: The current is above the permissible limit.
B078
Current I2 Load Error
Current output 2: The current loop has been interrupted (open circuit) or
the load is too high: Deactivate or short-circuit unused current outputs.
B079
Current I2 Parameter
Current output 2: Check the configuration.
B080
Current I3 Span
Current output 3: The selected span is too small/too big:
Parameter Setting
Inputs and Outputs
Current Outputs
Current Output I3
Check start/end.
B081
Current I3 < 0/4 mA
Current output 3: The current is below the permissible limit.
B082
Current I3 > 20 mA
Current output 3: The current is above the permissible limit.
B083
Current I3 Load Error
Current output 3: The current loop has been interrupted (open circuit) or
the load is too high: Deactivate or short-circuit unused current outputs.
B084
Current I3 Parameter
Current output 3: Check the configuration.
B085
Current I4 Span
Current output 4: The selected span is too small/too big:
Parameter Setting
Inputs and Outputs
Current Outputs
Current Output I4
Check start/end.
B086
Current I4 < 0/4 mA
Current output 4: The current is below the permissible limit.
B087
Current I4 > 20mA
Current output 4: The current is above the permissible limit.
B088
Current I4 Load Error
Current output 4: The current loop has been interrupted (open circuit) or
the load is too high: Deactivate or short-circuit unused current outputs.
B089
Current I4 Parameter
Current output 4: Check the configuration.