10 Troubleshooting
10.1 Malfunction States
Messages and errors are displayed with the corresponding NAMUR icon and the measurement display
of the corresponding channel changes color.
The message is logged, with time and date, in the logbook.
If messages are set to current outputs or relay contacts, these are activated after the user-defined
delay has expired.
Displaying Messages
Go to the Diagnostics menu if the “failure”
, “maintenance required”
, or “out of specification”
icons are flashing on the display.
Menu Selection
Diagnostics
Message List
All active messages are displayed in the “Message List” menu item with the following information:
Error number, type (failure, maintenance required, out of specification), channel, message text.
You can scroll forward and backward using the
up/down arrow keys
.
Message List
F031
Back
P113
Sensor Operating Time
No Module Connected
II
The error message disappears from the display around 2 s after the error is cleared.