2 Product
2.1 Design and Function
Stratos Multi is designed as follows:
• The 1- and 2-channel transmitter permits a free combination of the pH/ORP process variables,
conductivity (2-/4-electrode sensors, toroidal sensors), and oxygen, and can, for example,
simultaneously measure pH values and conductivity.
• The device has four current outputs (active/passive).
• Add-on product options can be enabled by entering a transaction number (TAN).
• The color TFT display has white backlighting and provides clear information on errors during
parameter setting or measurement in accordance with NAMUR recommendations.
• The user interface is multi-lingual with full-text menu navigation presently available in
the following languages: German, English, French, Italian, Spanish, Portuguese, and Chinese.
Function
TAN option
1 measuring channel
2 current outputs
Second measuring channel via add-on measuring module
3rd and 4th current outputs
X
Multiparameter: Process variables freely selectable from pH, ORP, oxygen,
conductivity (contacting/inductive)
Door contact
HART
X
3 freely assignable relay contacts
for NAMUR messages (failure, maintenance required, out of specification, function check),
setpoint alarm relay, controller, rinse contact parameter set, USP (for conductivity), Sensoface
Function check input (HOLD)
2 control Inputs
Flow measurement
Current Input
X