Stratos Multi E401N
67
Adjusting Sensor Monitoring
01.
Open
Sensor Data
Sensor Monitoring Details
.
02.
Open a sensor parameter, e.g.,
Slope
.
03.
Set
Monitoring
of the slope to automatic or individual.
04.
If you select “Individual”: The nominal slope and the min/max limits can be entered.
05.
In the
Message
menu item, select whether and how an exceeded limit is to be displayed:
Off
No message, but the parameter is still shown on the sensor diagram.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Mainte-
nance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
06.
Set the sensor monitoring details for other sensor data, e.g. zero point, Sensocheck, response time,
sensor wear, or sensor operating time.
07.
Using the
left softkey: Back
, confirm the sensor monitoring settings and adjust further parameters.
or
Using the
right softkey: Back to Meas.
, confirm the sensor monitoring settings and end the func-
tion check (HOLD).