6.6.4 Calibration Mode: Data Entry
Entry of values for slope and zero point of the sensor, related to 25 °C/77 °F, 1013 mbar/14.69 psi.
Slope = sensor current at 100% atmospheric oxygen, 25 °C/77 °F, 1013 mbar/14.69 psi
Calibration Procedure
Calibration
[I] [II] … Oxy
When you access calibration, the calibration values selected in the calibration presettings are used.
These settings can still be changed in the Calibration menu. If you do not want to perform calibration,
use the left
softkey
to go back one level and abort the calibration.
01.
Select “Data Entry” calibration mode and press
enter
to confirm.
02.
Remove the sensor and install the premeasured sensor.
03.
Continue with the
right softkey: Next
.
04.
Enter the zero point and slope measured values, and press
enter
to confirm.
Result
✔
By pressing the
right softkey: Adjust
, the calibration values obtained during calibration for
calculation of the process variables are applied to the device. When using a Memosens sensors,
the calibration values are saved in the sensor.