Usage: Rinse Contact
Please note when configuring the “Rinse Contact” function
• The “function check” (HOLD) mode (e.g. during parameter setting) delays the execution of the
“rinse contact” function.
• Up to 3 rinse functions (contacts K1 ... K3) can be configured independently of each other.
• The individual rinse functions are not synchronized with each other.
Configuring the Rinse Contact
01.
Open
Inputs and Outputs
Relay Contacts
.
02.
Select the desired contact with the
up/down arrow keys
and
enter
.
03.
Usage: Rinse Contact
04.
Select
contact type
(e.g. “N/O”).
05.
Specify
rinsing interval
.
06.
Specify
rinse duration
.
07.
Specify
rinse/meas. lead time.
Note:
Function check (HOLD) is active during the defined “Lead Time...”
08.
Logbook entry
: off/on
Usage: USP Output
Can be activated when using a conductivity sensor and the USP function
01.
Open
Inputs and Outputs
Relay Contacts
.
02.
Select the desired contact with the
up/down arrow keys
and
enter
.
03.
Usage: USP Output
04.
Assign
USP channel
.
05.
Set the contact parameters.
Usage: Sensoface
Sensoface messages can be output via a relay contact.
If using two sensors, the Sensoface messages can be assigned to different contacts:
01.
Open
Inputs and Outputs
Relay Contacts
.
02.
Select the desired contact with the
up/down arrow keys
and
enter
.
03.
Usage: Sensoface Channel
04.
Select
channel
.
Contact K1 (Admin.)
Channel
Back
Usage
Contact Type
ON Delay
OFF Delay
Sensoface Channel
Memosens pH
N/O..
I
Memosens pH
Memosens Cond
II
0 s
0 s
05.
Set the contact parameters.
06.
Using the
left softkey: Back
, go to the
Relay Contacts
menu and configure the second contact.
See also