Stratos Multi E401N
145
Error no. Type Message text
Notes/solution
E036
PASSAGE LO_LO
Value below configured monitoring limit
E037
PASSAGE LO
Value below configured monitoring limit
E038
PASSAGE HI
Value above configured monitoring limit
E039
PASSAGE HI_HI
Value above configured monitoring limit
E045
REJECTION Range
Rejection. Min/max device limits exceeded:
- Check both conductivity values.
E046
REJECTION LO_LO
Value below configured monitoring limit
E047
REJECTION LO
Value below configured monitoring limit
E048
REJECTION HI
Value above configured monitoring limit
E049
REJECTION HI_HI
Value above configured monitoring limit
E050
DEVIATION Range
Deviation. Min/max device limits exceeded:
- Check both conductivity values.
E051
DEVIATION LO_LO
Value below configured monitoring limit
E052
DEVIATION LO
Value below configured monitoring limit
E053
DEVIATION HI
Value above configured monitoring limit
E054
DEVIATION HI_HI
Value above configured monitoring limit
E055
Remaining Capacity
Range
The remaining capacity of the ion exchanger cannot be calculated.
E056
Degassed
Conductivity
Min/max device limits exceeded:
- Check both conductivity values.
E057
Ion Exchanger
Remaining Capacity
Remaining capacity of the ion exchanger < 20 %:
Check ion exchanger; as required, replace the filter or the ion exchanger.
Remaining capacity of the ion exchanger 0 %:
Replace the ion exchanger.
Replacement of the ion exchanger must be confirmed in the Maintenance
menu:
Maintenance
[CI] [CII] Cond-Cond Calculation
E060
pH Range
pH range outside the permitted range of the VGB guideline:
- Check both conductivity values.
- Check choice of alkalizing agent.
- Check ion exchanger.
- Check both sensors/cables.
E061
pH LO_LO
Value below configured monitoring limit
E062
pH LO
Value below configured monitoring limit
E063
pH HI
Value above configured monitoring limit
E064
pH HI_HI
Value above configured monitoring limit
E200
Param. Calculation
Block
Check the configuration:
Parameter Setting
System Control
Calculation Blocks