10.2 Error Messages
Error type
Failure
Maintenance Required
Out of Specification
Info
Info text, appears directly in the relevant menu
par
User-defined error type: failure or maintenance required
General Error Messages
Error
Notes/solution
Display is blank
Press any key to wake the display following a possible auto-off.
Check the voltage supply.
No measurement, no error message
Check the sensor connection/install the module properly.
Configure the measurement display.
Error no. Type Message text
Notes/solution
F008
Adjustment Data
Error in the adjustment data:
Switch device off (approx. 10 s).
If the message persists, send in the device.
F009
Firmware Error
Error in the firmware:
Switch device off (approx. 10 s).
Reload the firmware.
Firmware Update (FW-E106), p. 213
If the message persists, send in the device.
F029
No Sensor Connected The sensor is not recognized:
- Check connections.
- Check cables. Replace as required.
- Check sensor. Replace as required.
F030
Wrong Sensor
Connected
The connected digital sensor does not match the configuration:
- Connect correct sensor.
- Adjust process variable.
Sensor Selection [I] [II], p. 63
F031
No Module Con-
nected
No module is recognized. Possible causes:
- No module installed.
- Wrong module selected.
- Module faulty.
Install the module properly and select it in the parameter settings.
F033
Info
Sensor Removed
The sensor cannot be found. Possible causes:
- Sensor was removed.
- Connections/cables faulty.
Connect an appropriate sensor and adjust the configuration as
necessary.
F038
Sensor Defective
Sensor is faulty. Replace the sensor.
F191
Info
Meas. Recorder
Inconsistent
The measurement recorder’s memory is defective and cannot be read:
Deleting the measurement recorder’s data.
F200
Configuration Data
Loss
Data error in the configuration: Reset to factory settings and fully
reconfigure.
F202
System Failure
Internal system error: Switch device off (approx. 10 s).
If the message persists, send in the device.
F203
Inconsistent
Parameter Setting
The configuration of the measuring channel operating mode is
inconsistent: Check and correct the configuration.
F210
Device Diagnostics:
Self Test
Error on the self test (RAM): Switch device off (approx. 10 s).
If the message persists, send in the device.