Stratos Multi E401N
74
Note:
If measurements are generally taken at high temperatures (> 55 °C/131 °F), the counters should
be switched off.
When a CIP/SIP counter is switched on, a maximum number of cycles can be entered. As soon as the
counter has reached the specified value, a maintenance required message is triggered and the NAMUR
icon
is displayed.
Note:
A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle
is complete.
Note:
With Memosens sensors, an entry is also made in the sensor.
5.8.2 Calibration Presettings
The calibration presettings can be defined in the parameter settings or adjusted directly in the Calibra-
tion menu prior to the calibration.
Calibration mode
: Presetting of the calibration mode, e.g., ORP data entry, ORP adjustment, ORP
check, temperature
Calibration timer
: When a preset calibration interval expires, the calibration timer generates a mes-
sage text to indicate the need for calibration. If “Auto” is selected, the interval is set to 168 h. If “Individ-
ual” is selected, you can select a custom interval.
Note:
If Sensoface is enabled, the smiley is “sad” as soon as the interval has expired. You can configure
the current outputs so that a Sensoface message generates a 22-mA error signal.
ORP check:
Settings for test period in seconds and test difference in millivolts
Cal Presettings (Admin.)
Back
Calibration Mode
Drift Check
Calibration Timer
ORP Check
ORP Data Entry
Standard
II
Data Entry
ORP Data Entry
ORP Adjustment
ORP Check
Temperature
Settings can be changed in the
Cal Presettings
submenu:
Parameter Setting
[I] [II] [Sensor]
Cal Presettings
5.8.3 Delta Function
Note:
If the delta function is enabled, “∆” is shown on the display in measuring mode.
If a delta value is specified, the measuring system calculates the difference
output value = measured value – delta value
The delta value can be set using the “+” or “-” signs. If using a negative sign, the delta value is added to
the measured value.
The delta value is adjusted in the
Delta Function
submenu:
Parameter Setting
[I] [II] [Sensor]
Delta Function
All outputs are controlled by the output value; the displays show the output value.
Note:
If the delta function and TC correction are enabled at the same time, the TC correction is carried
out first and the delta value is then deducted.