Stratos Multi E401N
90
Selecting a Second Memosens Oxygen Sensor
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [II]
Selection of a second Memosens oxygen sensor connected to the MK-MS 095N measuring module:
Module:
MK-MS
Parameter:
Oxygen
Mode:
Memosens
Functionality:
Amperometric
Adjustable parameters for Memosens oxygen sensors
Parameter Setting
[I] [II] Memosens Oxy
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
Measure in
Liquids, gases
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Option to enter individual limits for monitoring separate pa-
rameters.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance required
messages.
Option to enter individual limits before a message for response
time, sensor wear, sensor operating time, SIP counter is trig-
gered.
Cal Presettings
Presetting of the calibration mode and calibration timer.
Calibration Presettings, p. 95
Pressure Correction
Manual input of pressure during measurement and calibration.
With TAN option FW-E051: automatic pressure correction by external pressure transmit-
ter.
Salinity Correction
Salinity, chlorinity, conductivity
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
Selecting an SE740 Optical Oxygen Sensor (LDO)
Parameter Setting
Sensor Selection [I] [II]
Sensor Selection [I]
Selection of the SE740 optical oxygen sensor connected to the RS-485 interface (terminals 1 … 6):
Parameter:
Auto or oxygen
Mode:
Other digital
Functionality:
SE740
Note:
The voltage at terminal 6 (Power Out) is automatically set to 14 V when using the SE740 optical
oxygen sensor. The Power Out menu item is not available.
Adjustable parameters for the SE740 optical oxygen sensor
Parameter Setting
[I] Digital Oxy
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
Measure in
Liquids, gases
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Select whether a failure or maintenance required message is
triggered if the range of the Stern-Volmer constant and phase
angle is exceeded.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance required
messages.
Option to enter individual limits before a message for sensor
wear, sensor operating time, CIP counter, O
2
measurement with
CIP/SIP, autoclaving counter is triggered.