6.5.6 Calibration Mode: Zero Point
Zero point correction
Calibration Procedure
Calibration
[I] [II] … Condl
When you access calibration, the calibration values selected in the calibration presettings are used.
These settings can still be changed in the Calibration menu. If you do not want to perform calibration,
use the left
softkey
to go back one level and abort the calibration.
01.
Select “Zero Point” calibration mode.
02.
Take the sensor out of the medium, rinse it in deionized water, and dry it. The sensor should be
dry, since zero calibration is performed in air.
03.
Press the
right softkey: Next
.
✓
Zero point correction is carried out. The permissible zero deviation is model-dependent; on the
SE 670 sensor, for example, it is ± 0.050 mS/cm.
04.
Press the
right softkey: Next
.
Result
✔
By pressing the
right softkey: Adjust
, the calibration values obtained during calibration for
calculation of the process variables are applied to the device. When using a Memosens sensors,
the calibration values are saved in the sensor.
Valid for toroidal conductivity with Memosens sensors:
The calibration values are displayed following successful zero calibration.
05.
Press the
right softkey: Next
.
✓
The following message is shown: Adjustment Successful.
6.5.7 Calibration Mode: Installation Factor
When using a Memosens sensor in a tight space, the installation factor is entered.
Calibration Procedure
Calibration
[I] [II] … Condl
When you access calibration, the calibration values selected in the calibration presettings are used.
These settings can still be changed in the Calibration menu. If you do not want to perform calibration,
use the left
softkey
to go back one level and abort the calibration.
01.
Select “Install. Factor“ calibration mode and press
enter
to confirm.
02.
Enter the installation factor.
03.
Next with the
right softkey: Save
.
✓
The following message is shown: Adjustment Successful.