Stratos Multi E401N
141
Error no. Type Message text
Notes/solution
P124
Sensor Date
The date of the sensor is implausible.
e.g., calibration data “from the future”: Check and, as necessary, adjust the
configuration.
P201
Cal: Temperature
The calibration temperature is impermissible: Check the calibration temper-
ature. Note the information in the Calibration chapter.
Process Variable pH Calibration/Adjustment, p. 99
P202
Info
Cal: Buffer Unknown
Calibration error with automatic Calimatic calibration: Buffer not recog-
nized. Possible causes: Incorrect buffer set selected. Buffer corrupted. Sen-
sor faulty: Check calibration. Note the information in the Calibration chap-
ter.
Calibration Mode: Calimatic, p. 101
P203
Info
Cal: Identical Buffers
Calibration error with automatic Calimatic calibration: The same buffer was
used. The sensor or the sensor cable may be faulty.
P204
Info
Cal: Buffers Inter-
changed
Calibration error during manual calibration: Buffer order differs from speci-
fied order. Repeat calibration and adhere to order.
Calibration Mode: Manual, p. 103
P205
Info
Cal: Sensor Unstable
The drift criterion was not adhered to during calibration. Possible causes:
improper calibration, sensor cable/connection faulty, sensor worn.
Check the sensor and calibration and repeat as required. Otherwise, replace
the sensor.
P206
Cal: Slope
Slope out of permissible range: Repeat calibration/adjustment or replace
sensor.
P207
Cal: Zero Point
Zero point out of permissible range: Repeat calibration/adjustment or re-
place sensor.
P208
Cal: Sensor Failure
Replace the sensor.
Calculation Block pH/pH
Error no. Type Message text
Notes/solution
A001
Configuration Data
Loss
Data error in the configuration:
Fully reconfigure the device.
A010
pH Diff Range
pH value difference. Max/min device limits exceeded:
- Check both pH values.
- Check sensor/cable connections.
A011
pH Diff LO_LO
Value below configured monitoring limit
A012
pH Diff LO
Value below configured monitoring limit
A013
pH Diff HI
Value above configured monitoring limit
A014
pH Diff HI_HI
Value above configured monitoring limit
A015
Temperature Diff
Range
Temperature value difference. Max/min device limits exceeded:
- Check both temperature values.
- Check sensor/cable connections.
A016
Temperature Diff
LO_LO
Value below configured monitoring limit
A017
Temperature Diff LO
Value below configured monitoring limit
A018
Temperature Diff HI
Value above configured monitoring limit
A019
Temperature Diff
HI_HI
Value above configured monitoring limit
A020
ORP Diff Range
ORP value difference. Max/min device limits exceeded:
- Check both ORP values.
- Check sensor/cable connections.
A021
ORP Diff LO_LO
Value below configured monitoring limit
A022
ORP Diff LO
Value below configured monitoring limit