Error no. Type Message text
Notes/solution
D061
Neutral Sensoface
Zero Point
- Readjust sensor.
- Check/refill electrolyte.
- SE740: Replace membrane cap.
- Replace sensor.
D062
par
Sad Sensoface
Sensocheck
- Readjust configured sensor.
- Replace sensor.
D063
Sad Sensoface
Response Time
- Check/refill electrolyte.
- SE740: Replace membrane cap.
- Replace sensor.
D064
Sad Sensoface
Cal Timer
Calibration timer has expired:
- Check the calibration timer settings.
- Carry out calibration/adjustment.
D070
par
Sad Sensoface
Wear
Sensor is worn (100 %):
- Readjust sensor.
- Check/refill electrolyte.
- Replace sensor.
D071
par
Sad Sensoface
Membrane
For SE740 optical oxygen sensor only: Replace membrane cap.
D080
Sensor Current Range - Check polarization voltage:
Parameter Setting
[I] [II] … Oxy
Sensor Data
- Refill electrolyte.
- Recalibrate/readjust.
D081
O2 Measurement OFF
(Temp)
For SE740 optical oxygen sensor only: The oxygen measurement was auto-
matically stopped because the temperatures are too high.
D110
CIP Counter
Configured number of CIP cycles exceeded: As required, recalibrate/read-
just sensor or replace.
D111
SIP Counter
Configured number of SIP cycles exceeded: As required, recalibrate/read-
just sensor or replace.
D112
Autoclaving Counter
Configured number of autoclaving cycles exceeded: As required, recali-
brate/readjust sensor or replace.
D113
Sensor Operating
Time
Sensor operating time exceeded.
Replace the sensor.
D114
Membrane Body Re-
placement
Replacement of the membrane body required.
Replacement of the membrane body must be confirmed in the Mainte-
nance menu.
Maintenance Functions for Oxygen Sensors, p. 134
D120
Wrong Sensor
If sensor check is active: An impermissible sensor is connected to the
device.
- Replace sensor.
- Change the process variable.
D121
Sensor Error
(Factory Settings)
The digital sensor indicates an error. The sensor no longer works properly.
Replace the sensor.
D122
Sensor Memory (Cal
Data)
The digital sensor indicates an error. The calibration data are defective:
Recalibrate/readjust sensor.
D123
New Sensor,
Adjustment Required
A new digital sensor was connected. It must now be adjusted.
D124
Sensor Date
The date of the sensor is implausible.
e.g., calibration data “from the future”: Check and, as necessary, adjust the
configuration.
D200
Temp O2 Conc/Sat
The temperature is outside the valid range for oxygen concentration/
saturation.