Stratos Multi E401N
94
O
2
Measurement with CIP/SIP
When using the SE740 optical oxygen sensor, the temperature can be monitored during the CIP/SIP
process. When the preset cut-off temperature is exceeded, the sensor automatically stops the oxygen
measurement to extend the life of the membrane. The measured oxygen value is frozen,
the temperature continues to be measured.
01.
Open
Sensor Monitoring Details
O2 Measurement with CIP/SIP
.
02.
Monitoring
: Automatic or individual
03.
If you select “Individual”: Enter the maximum cut-off temperature.
04.
In the
Message
menu item, select whether and how an exceeded limit is to be displayed:
Off
No message, but the parameter is still shown on the sensor diagram.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Maintenance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Settings can be changed in the
Sensor Monitoring Details
submenu:
Parameter Setting
[I] [II] ... Oxy
Sensor Data
Sensor Monitoring Details
Autoclaving Counter
An autoclaving counter is available for the SE740 oxygen sensor.
When the autoclaving counter is switched on, a maximum number of cycles can be entered. As soon
as the autoclaving counter has reached the specified value, a maintenance required message is
triggered and the NAMUR icon
is displayed.
After each autoclaving process, you must manually increment the autoclaving counter in the
Maintenance menu.
Maintenance
[I] Digital Oxy
Sensor Monitor
Autoclaving Cycles