Stratos Multi E401N
152
Sensoface messages can be assigned to a relay contact:
Parameter Setting
Inputs and Outputs
Relay Contacts
Contact K…
Usage
If you select
Sensoface
, all Sensoface messages are output via the selected contact.
If you select
Sensoface (Channel)
, you can output the Sensoface messages from a specific channel via
the selected contact.
Sensoface Criteria
Conductivity (Contacting)
Sensoface
Cell constant
2-/4-electrode sensors
Memosens
Happy
0.005 cm
-1
… 19.9999 cm
-1
0.5 x nom. cell constant … 2 x nom. cell constant
Sad
< 0.005 cm
-1
or
> 19.9999 cm
-1
< 0.5 x nom. cell constant or
> 2 x nom. cell constant
Conductivity (Inductive)
Sensoface
Cell factor
Zero point
Analog sensors
Memosens
Happy
0.1 cm
-1
… 19.9999 cm
-1
0.5 x nom. cell factor … 2 x nom. cell
factor
-0.25 mS … 0.25 mS
Sad
< 0.1 cm
-1
or
> 19.9999 cm
-1
< 0.5 x nom. cell factor or
> 2 x nom. cell factor
< -0.25 mS or
> 0.25 mS
Note:
The worsening of a Sensoface criterion leads to the devaluation of the Sensoface indicator
(Smiley gets "sad"). An improvement of the Sensoface indicator can only take place after calibration or
removal of the sensor defect.
Sensocheck
Process variable
Sensocheck function
pH
Automatic monitoring of glass and reference electrode
Oxygen
Monitoring membrane/electrolyte
Conductivity
Information on sensor condition
Enable/Disable Sensocheck
Sensocheck is enabled and disabled in the
Sensor Data
submenu:
With Memosens:
Parameter Setting
[I] [II] Memosens …
Sensor Data
Sensor Monitoring Details
Sensocheck
You can enable or disable Sensocheck in the
Monitoring
menu item.
In the
Message
menu item, select whether a Sensocheck message is output as a failure or
maintenance required message.
With analog sensors:
Parameter Setting
[I] [II] [Sensor]
Sensor Data
Sensocheck
In the
Sensocheck
menu item, you can disable Sensocheck or choose to output a Sensocheck
message as a failure or maintenance required message.