Stratos Multi E401N
148
Error no. Type Message text
Notes/solution
T205
Info
Cal: Sensor Unstable
The drift criterion was not adhered to during calibration. Possible causes:
improper calibration, sensor cable/connection faulty, sensor worn.
Check the sensor and calibration and repeat as required. Otherwise,
replace the sensor.
Oxygen
Error no. Type Message text
Notes/solution
D001
Configuration Data
Loss
Data error in the configuration:
Fully reconfigure the device.
D008
Adjustment Data
Error in the adjustment data:
Switch device off (approx. 10 s).
If the message persists, send in the device.
D009
Firmware Error
Error in the firmware:
Switch device off (approx. 10 s).
Reload the firmware.
Firmware Update (FW-E106), p. 213
D010
Saturation %Air
Range
Max/min limits of range exceeded or sensor not/incorrectly connected, ca-
ble faulty.
D011
Saturation %Air
LO_LO
Value below configured monitoring limit
D012
Saturation %Air LO
Value below configured monitoring limit
D013
Saturation %Air HI
Value above configured monitoring limit
D014
Saturation %Air HI_HI Value above configured monitoring limit
D015
Temperature Range
Max/min limits of range exceeded or sensor not/incorrectly connected, ca-
ble faulty.
D016
Temperature LO_LO
Value below configured monitoring limit
D017
Temperature LO
Value below configured monitoring limit
D018
Temperature HI
Value above configured monitoring limit
D019
Temperature HI_HI
Value above configured monitoring limit
D020
Concentration Range
Max/min limits of range exceeded, sensor not/incorrectly connected,
cable faulty.
D021
Concentration LO_LO Value below configured monitoring limit
D022
Concentration LO
Value below configured monitoring limit
D023
Concentration HI
Value above configured monitoring limit
D024
Concentration HI_HI
Value above configured monitoring limit
D025
Partial Pressure Range Max/min limits of range exceeded, sensor not/incorrectly connected,
cable faulty.
D045
Saturation %O2
Range
Max/min limits of range exceeded, sensor not/incorrectly connected,
cable faulty.
D046
Saturation %O2
LO_LO
Value below configured monitoring limit
D047
Saturation %O2 LO
Value below configured monitoring limit
D048
Saturation %O2 HI
Value above configured monitoring limit
D049
Saturation %O2 HI_HI Value above configured monitoring limit
D060
Sad Sensoface
Slope/
- Readjust sensor.
- Check/refill electrolyte.
- SE740: Replace membrane cap.
- Replace sensor.