background image

5-2-55, Minamitsumori, Nishinari-ku, Osaka 557-0063 JAPAN
Phone: +81(6)6659-8201  Fax: +81(6)6659-8510  E-mail: [email protected]

EM-3561-CE  P. 1 / 3

INSTRUCTION MANUAL

CT TRANSMITTER

MODEL

KC/KCE

BEFORE USE ....

Thank you for choosing M-System.  Before use, please check 
contents of the package you received as outlined below.
If you have any problems or questions with the product, 
please contact M-System’s Sales Office or representatives.

 

PACKAGE INCLUDES:

Signal conditioner
(body + base CT protector) ...................................(1)

 

MODEL NO.

Confirm Model No. marking on the product to be exactly 
what you ordered.

 

INSTRUCTION MANUAL

This manual describes necessary points of caution when 
you use this product, including installation, connection and 
basic maintenance procedures.

POINTS OF CAUTION

 

CONFORMITY WITH EU DIRECTIVES

• This equipment is suitable for Pollution Degree 2 and 

Installation Category II (transient voltage 2500V).  Re-
inforced insulation (signal input or output to auxiliary 
power input: 300V) and basic insulation (signal input to 
output: 300V) are maintained.  Prior to installation, check 
that the insulation class of this unit satisfies the system 
requirements.

• Altitude up to 2000 meters.
• The equipment must be mounted inside a panel.
• The equipment must be installed such that appropriate 

clearance and creepage distances are maintained to con-
form to CE requirements.  Failure to observe these re-
quirements may invalidate the CE conformance.

• The actual installation environments such as panel con-

figurations, connected devices, connected wires, may af-
fect the protection level of this unit when it is integrated 
in a panel system.  The user may have to review the CE 
requirements in regard to the whole system and employ 
additional protective measures to ensure the CE conform-
ity.

• Install lightning surge protectors for those wires connect-

ed to remote locations.

 

AUXILIARY POWER INPUT RATING & OPERATIONAL RANGE

• Locate the auxiliary power input rating marked on the 

product and confirm its operational range as indicated 
below:

  

AC power

: Rating -15/+10%, 50/60 Hz, approx. 2VA

  

DC power

: Rating ±10%, approx. 2W

 

GENERAL PRECAUTIONS

• Before you remove the unit from its base socket or mount 

it, turn off the power supply and input signal for safety.

 

CT PROTECTOR

• Temperature of the CT Protector may rise if the signal 

conditioner body is left unplugged from the base socket.  
Be careful not to get burned when you need to touch it.

 

ENVIRONMENT

• Indoor use.
• When heavy dust or metal particles are present in the 

air, install the unit inside proper housing with sufficient 
ventilation.

• Do not install the unit where it is subjected to continuous 

vibration.  Do not subject the unit to physical impact.

• Environmental temperature must be within -10 to +55°C 

(14 to 131°F) with relative humidity within 30 to 85% RH 
in order to ensure adequate life span and operation.

 

WIRING

• Do not install cables close to noise sources (relay drive 

cable, high frequency line, etc.).

• Do not bind these cables together with those in which 

noises are present.  Do not install them in the same duct.

 

AND ....

• The unit is designed to function as soon as power is  

supplied, however, a warm up for 10 minutes is required 
for satisfying complete performance described in the data 
sheet.

COMPONENT IDENTIFICATION

CT Protector

K

-UNIT

Body

Base Socket

Specifications

Zero Adj.

Span Adj.

Отзывы: