Keysight E8357D/67D & E8663D PSG User’s Guide 323
Troubleshooting
Data Storage Problems
Data Storage Problems
Registers With Previously Stored Instrument States are Empty
The save/recall registers are backed–up by a battery when line power to the
signal generator is not connected. The battery may need to be replaced.
To verify that the battery has failed:
1.
Turn off line power to the signal generator.
2.
Unplug the signal generator from line power.
3.
Plug in the signal generator.
4.
Turn on the signal generator.
5.
Observe the display for error messages.
If either error message
−
311 or
−
700 is stored in the error message queue,
the signal generator’s battery has failed.
6.
Refer to the Keysight PSG Signal Generators Service Guide for battery
replacement instructions.
Saved Instrument State, but Register is Empty or Contains Wrong
State
If you select a register number greater than 99, the signal generator
automatically selects register 99 to save the instrument state.
If the register number you intended to use is empty or contains the wrong
instrument state, recall register 99:
Press
Recall
>
99
>
Enter
.
The lost instrument state may be saved there.
Содержание E8663D
Страница 4: ...4...
Страница 58: ...44 Keysight E8357D 67D E8663D PSG User s Guide Signal Generator Overview Rear Panel...
Страница 186: ...172 Keysight E8357D 67D E8663D PSG User s Guide Optimizing Performance Wideband IQ FM Mode 3 2 GHz Option 018...
Страница 198: ...184 Keysight E8357D 67D E8663D PSG User s Guide Analog Modulation Configuring the LF Output Option UNT...
Страница 212: ...198 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Working with Filters...
Страница 226: ...212 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Configuring Hardware...
Страница 286: ...272 Keysight E8357D 67D E8663D PSG User s Guide AWGN Waveform Generator Configuring the AWGN Generator...
Страница 326: ...312 Keysight E8357D 67D E8663D PSG User s Guide Peripheral Devices Millimeter Wave Source Modules...
Страница 346: ...332 Keysight E8357D 67D E8663D PSG User s Guide Troubleshooting Returning a Signal Generator to Keysight Technologies...