Keysight E8357D/67D & E8663D PSG User’s Guide 37
Signal Generator Overview
Rear Panel
For Option UNR/UNX or instruments with serial prefixes > US4805/MY4805,
this BNC connector accepts a signal with a nominal input level of 5
±
5 dBm.
The external frequency reference must be 10 MHz, within
±
1 ppm.
The nominal input impedance is 50 ohms with a damage level of
≥
10 dBm.
21. LAN
This LAN interface allows ethernet local area network communication through
a 10Base–T LAN cable. The yellow LED on the interface illuminates when data
transmission (transfer/receive) is present. The green LED illuminates when
there is a delay in data transmission or no data transmission is present.
22. 10 MHz OUT
This female BNC connector outputs a nominal signal level of >
+
4 dBm and has
an output impedance of 50 ohms. The accuracy is determined by the timebase
used.
23. STOP SWEEP IN/OUT
This female BNC connector (Option 007 only) provides an open–collector,
TTL–compatible input/output signal that is used during ramp sweep operation.
It provides low–level (nominally 0 V) output during sweep retrace and
band–cross intervals. It provides high–level (nom5 V) output during the
forward portion of the sweep. Sweep stops when this input/output connector is
grounded externally. When operating as an input, the nominal impedance for
this connector is less than 10 ohms. When operating as an output, the nominal
impedance is approximately 4.2 kohms.
24. BASEBAND GEN CLK IN
This female BNC connector accepts a sine or square wave PECL clock input
with a frequency range of 200 MHz to 400 MHz, resulting in sample rates of
50 MSa/s to 100 MSa/s. The recommended input level is approximately 1
Vpeak–to–peak for a square wave and 1 dBm to 6 dBm for a sine wave. This
allows baseband generators from multiple signal sources to run off the same
clock.
25. Z–AXIS BLANK/MKRS
This female BNC connector (Option 007 only) supplies a +5 V (nominal) level
during retrace and band–switch intervals of a step, list, or ramp sweep. During
ramp sweep, this female BNC connector supplies a –5 V (nominal) level when
the RF frequency is at a marker frequency and intensity marker mode is on.
This signal is derived from an operational amplifier output so the load
impedance should be greater than or equal to 5 kohms. This connection is
most commonly used to interface with an Keysight 8757D Scalar Network
Analyzer.
Содержание E8663D
Страница 4: ...4...
Страница 58: ...44 Keysight E8357D 67D E8663D PSG User s Guide Signal Generator Overview Rear Panel...
Страница 186: ...172 Keysight E8357D 67D E8663D PSG User s Guide Optimizing Performance Wideband IQ FM Mode 3 2 GHz Option 018...
Страница 198: ...184 Keysight E8357D 67D E8663D PSG User s Guide Analog Modulation Configuring the LF Output Option UNT...
Страница 212: ...198 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Working with Filters...
Страница 226: ...212 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Configuring Hardware...
Страница 286: ...272 Keysight E8357D 67D E8663D PSG User s Guide AWGN Waveform Generator Configuring the AWGN Generator...
Страница 326: ...312 Keysight E8357D 67D E8663D PSG User s Guide Peripheral Devices Millimeter Wave Source Modules...
Страница 346: ...332 Keysight E8357D 67D E8663D PSG User s Guide Troubleshooting Returning a Signal Generator to Keysight Technologies...