Keysight E8357D/67D & E8663D PSG User’s Guide 155
Optimizing Performance
Creating and Applying User Flatness Correction
NOTE
A power meter timeout may occur at low power levels. If a power meter
timeout error message appears, increase the timeout value by pressing
Amplitude
>
More (1 of 2)
>
User Flatness > More (1 of 2) > Meter Timeout
.
2.
Press
Done
.
Pressing the
Done
softkey loads the amplitude correction values into the
user flatness correction array.
If desired, press
Configure Cal Array
.
This opens the user flatness correction array, where you can view the
stored amplitude correction values. The user flatness correction array title
displays
User Flatness: (UNSTORED)
indicating that the current user
flatness correction array data has not been saved to the memory catalog.
Performing the User Flatness Correction Manually
If you are not using an Keysight E4416A/17A/18B/19B power meter, or if your
power meter does not have a GPIB interface, complete the steps in this section
and then continue with the user flatness correction tutorial.
1.
Press
More (1 of 2)
>
User Flatness
>
Configure Cal Array
.
This opens the User Flatness table editor and places the cursor over the
frequency value
(1 GHz) for row 1. The RF output changes to the frequency value of the
table row containing the cursor and
1.000 000 000 00
is displayed in
the
AMPLITUDE
area of the display.
2.
Observe and record the measured value from the power meter.
3.
Subtract the measured value from 0 dBm.
4.
Move the table cursor over the correction value in row 1.
5.
Press
Edit Item
> enter the difference value from step 3 >
dB
.
The signal generator adjusts the RF output amplitude based on the
correction value entered.
6.
Repeat steps 2 through 5 until the power meter reads 0 dBm.
7.
Use the down arrow key to place the cursor over the frequency value
for the next row. The RF output changes to the frequency value of the
table row containing the cursor, as shown in the
AMPLITUDE
area of the
display.
8.
Repeat steps 2 through 7 for every entry in the User Flatness table.
Содержание E8663D
Страница 4: ...4...
Страница 58: ...44 Keysight E8357D 67D E8663D PSG User s Guide Signal Generator Overview Rear Panel...
Страница 186: ...172 Keysight E8357D 67D E8663D PSG User s Guide Optimizing Performance Wideband IQ FM Mode 3 2 GHz Option 018...
Страница 198: ...184 Keysight E8357D 67D E8663D PSG User s Guide Analog Modulation Configuring the LF Output Option UNT...
Страница 212: ...198 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Working with Filters...
Страница 226: ...212 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Configuring Hardware...
Страница 286: ...272 Keysight E8357D 67D E8663D PSG User s Guide AWGN Waveform Generator Configuring the AWGN Generator...
Страница 326: ...312 Keysight E8357D 67D E8663D PSG User s Guide Peripheral Devices Millimeter Wave Source Modules...
Страница 346: ...332 Keysight E8357D 67D E8663D PSG User s Guide Troubleshooting Returning a Signal Generator to Keysight Technologies...