322 Keysight E8357D/67D & E8663D PSG User’s Guide
Troubleshooting
Sweep Problems
3.
Edit the dwell values if they are incorrect.
NOTE
The effective dwell time at the RF OUTPUT connector is the sum of the value
set for the dwell plus processing time, switching time, and settling time. This
additional time added to the dwell is generally a few milliseconds. The
TTL/CMOS output available at the TRIG OUT connector, however, is asserted
high only during the actual dwell time.
If the list dwell values are correct, continue to the next step.
4.
Observe if the
Dwell Type List Step
softkey is set to Step.
When Step is selected, the signal generator will sweep the list points
using the dwell time set for step sweep rather than the sweep list dwell
values.
To view the step sweep dwell time, follow these steps:
a.
Press
Configure Step Sweep
.
b.
Observe the value set for the
Step Dwell
softkey.
List Sweep Information is Missing from a Recalled Register
List sweep information is not stored as part of the instrument state in an
instrument state register. Only the current list sweep is available to the signal
generator. List sweep data can be stored in the instrument catalog. For
Содержание E8663D
Страница 4: ...4...
Страница 58: ...44 Keysight E8357D 67D E8663D PSG User s Guide Signal Generator Overview Rear Panel...
Страница 186: ...172 Keysight E8357D 67D E8663D PSG User s Guide Optimizing Performance Wideband IQ FM Mode 3 2 GHz Option 018...
Страница 198: ...184 Keysight E8357D 67D E8663D PSG User s Guide Analog Modulation Configuring the LF Output Option UNT...
Страница 212: ...198 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Working with Filters...
Страница 226: ...212 Keysight E8357D 67D E8663D PSG User s Guide Custom Arb Waveform Generator Configuring Hardware...
Страница 286: ...272 Keysight E8357D 67D E8663D PSG User s Guide AWGN Waveform Generator Configuring the AWGN Generator...
Страница 326: ...312 Keysight E8357D 67D E8663D PSG User s Guide Peripheral Devices Millimeter Wave Source Modules...
Страница 346: ...332 Keysight E8357D 67D E8663D PSG User s Guide Troubleshooting Returning a Signal Generator to Keysight Technologies...