4-8
KLP 031308
4. Click the
PRESS TO ACCEPT
button to accept the value. The status display reads
CMAX
.
Click the “+” button to increase and the “–“ to decrease the output current until the DVM
reads as close as possible to maximum rated current (Calculate current as follows:
I (Amperes) = DVM reading (Volts) / Shunt Resistance (Ohms).
5. Click the
PRESS TO ACCEPT
button to accept the value. The status display reads
OUTC
.
Connect to DVM (+) to Analog I/O Port J2, pin 7 and DVM (–) to pin 13. Click the “+” button
to increase and the “–“ to decrease the output current until the DVM reads as close as pos-
sible to 10V.
6. Click the
PRESS TO ACCEPT
button to accept the value. The status display will not change.
Click the
ANALOG REF
button on the calibration window to proceed to external calibration
(see PAR. 4.4.3).
4.4.3
ANALOG REFERENCE CALIBRATION (VXI
plug&play
DEMO)
During analog reference calibration the external analog reference voltages used to establish full
scale output current and voltage are defined. The factory default calibration of 10V programs full
scale output, however the following procedure allows any d-c value up to 10V to program full
scale, while a voltage proportionately less programs a corresponding decrease in the output.
NOTE: The full scale programming voltage has a 1:1 correspondence with the external read-
back; i.e., 10V full scale programming voltage results in a 10V full scale readback. If
the full scale programming voltage is changed to a value other than 10V, the full scale
readback voltage is changed accordingly.
1. Click the
ANALOG REF
button on the Calibration Window; the Analog Reference Calibration
Window (Figure 4-4) appears and the status display on the VXI main panel reads
INPV
.When the reference voltage is at the desired value, click the
PRESS TO ACCEPT
but-
ton to accept the value. When the calibration is saved, this analog voltage will be equivalent
to full scale output voltage. NOTE:
The full scale output voltage is determined by the vir-
tual settings in effect.
FIGURE 4-4. ANALOG REFERENCE CALIBRATION WINDOW
Содержание KLP 10-150-1.2K
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP031308 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 18: ...1 6 KLP 031308 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 19: ...KLP 031308 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 24: ......
Страница 27: ...KLP HV 031308 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......