KLP031308
3-15
NOTE: If a SYSTem:SECurity:IMMediate command is issued, the above relay control settings
are reset to the factory default:
FLT - NORM
.
3.2.13.4 QUICK BOOT
Enter
UTIL
menu (see PAR. 3.2.13, steps 1 through 3) and rotate either
VOLTAGE
or
CUR-
RENT
control until status display reads
QkBT
- Quick Boot. If Quick boot is ON, then the next
time the unit is powered up it will go directly to setpoint mode (status display reads
SET
, see
PAR. 3.2.7) without displaying
MODL
,
VIRT
, and
PROT
settings.
• The
DC VOLTS
display indicates whether quick boot is
ON
or
OFF
.
1. Tap the
DC OUTPUT
switch to change the setting. The
DC VOLTS
display begins to blink.
2. Rotate either
VOLTAGE
or
CURRENT
control until the desired quick boot state is displayed,
then tap the
DC OUTPUT
switch to save the setting; the blinking stops.
3. To exit, rotate either
VOLTAGE
or
CURRENT
control until the status display reads
RTN
,
then tap the
DC OUTPUT
switch. To exit the
UTIL
menu rotate either
VOLTAGE
or
CUR-
RENT
control until the status display reads
EXIT
, then tap the
DC OUTPUT
switch.
3.2.13.5 SYSTEM CONFIGURATION FUNCTIONS
The system configuration functions allow the user to determine 1) whether the Remote Inhibit
signal is an active TTL high or low and 2) how the unit recovers from a fault. The unit password
is required to change these functions.
1. Enter
UTIL
menu (see PAR. 3.2.13, steps 1 through 3) and rotate either
VOLTAGE
or
CUR-
RENT
control until status display reads
SYST
.
2. Tap the
DC OUTPUT
switch to enter the System Configuration Functions menu.
3.2.13.5.1 REMOTE INHIBIT CONFIGURATION
The Remote Inhibit signal REM_INH is applied to analog I/O port, pin 8 (see PAR’s. 2.7.9 and
3.4.1). This signal can be configured to be active (inhibit the output) for either a TTL high or low.
1. Enter
UTIL
menu (see PAR. 3.2.13, steps 1 through 3) and rotate either
VOLTAGE
or
CUR-
RENT
control until status display reads
SYST
, then tap
DC OUTPUT
switch. Rotate either
VOLTAGE
or
CURRENT
control until the status display reads
rINH
and the
DC VOLTS
dis-
play reads either
Hi
or
Lo
.
2. To change the remote inhibit setting, tap the
DC OUTPUT
switch. The status display shows
PASS
. Enter the unit password (see PAR. 3.2.5) and tap
DC OUTPUT
switch to accept the
number. When the password is accepted, the status display reads
rINH
and
DC VOLTS
dis-
play is blinking.
3. Rotate either
VOLTAGE
or
CURRENT
control until
DC VOLTS
display reads desired set-
ting, then tap
DC OUTPUT
switch to accept.
4. Tap
FUNCTION
switch to exit or rotate
VOLTAGE
or
CURRENT
control until status display
reads
RTN
, then tap
DC OUTPUT
switch to return to next highest menu.
Содержание KLP 10-150-1.2K
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP031308 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 18: ...1 6 KLP 031308 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 19: ...KLP 031308 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 24: ......
Страница 27: ...KLP HV 031308 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......