4-6
KLP 031308
7. Use the computer keyboard to enter the password in the text box displaying the word PASS-
WORD, then press the ENTER key.
8. If password is correct, the status display on the VXI main panel reads CAL and the buttons
on Calibration Screen (Figure 4-1) are enabled (highlighted). If the password is incorrect,
the password text box displays "WRONG #". Repeat step 7 to enter the correct password, or
click the
Calibrate Close
button on the VXI main panel to exit calibration.
NOTE: Calibrations may be done in any order.
9. Proceed to PAR. 4.4.1 for voltage calibration, PAR 4.4.2 for current calibration., PAR. 4.4.3
for external calibration or PAR. 4.4.4 to exit calibration
4.4.1
VOLTAGE CALIBRATION (VXI
plug&play
DRIVER DEMO)
During voltage calibration, the voltage, voltage readback and remote analog voltage readback
are calibrated.
1. Click the
VOLTAGE
button on the Calibration window to start voltage calibration: the Voltage
Calibration Window (Figure 4-2) appears. Minimum voltage is first; the status display (on the
VXI main panel) reads
V_0
. Monitor DVM and click the “+” button to increase and the “–“ to
decrease the output voltage until the DVM reads as close as possible to minimum rated
POSITIVE voltage.
FIGURE 4-2. VOLTAGE CALIBRATION WINDOW
2. Click the
PRESS TO ACCEPT
button to accept the value. The status display then reads
VMAX
. Monitor
DVM and click the “+” button to increase and the “–“ to decrease the output
voltage until the DVM reads as close as possible to maximum rated voltage.
Содержание KLP 10-150-1.2K
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP031308 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 18: ...1 6 KLP 031308 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 19: ...KLP 031308 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 24: ......
Страница 27: ...KLP HV 031308 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......