KLP 031308
4-7
3. Click the
PRESS TO ACCEPT
button to accept the value. The status display then reads
OUTV
. Connect
to DVM (+) to Analog I/O Port J2, pin 6 and DVM (–) to pin 13. Click the “+”
button to increase and the “–-“ button to decrease the output voltage until the DVM reads as
close as possible to 10V.
4. Click the
PRESS TO ACCEPT
button to accept the value. The status display will not change.
At the calibration window, click the
CURRENT
Button to proceed to current calibration (see
PAR. 4.4.2)
.
4.4.2
CURRENT CALIBRATION (VXI
plug&play
DEMO)
During current calibration, the current, current readback and analog current readback are cali-
brated.
1. Click the
CURRENT
button on the Calibration Window to start current calibration; the Cur-
rent Calibration Window (Figure 4-3) appears and the status display on the VXI main panel
reads
CMAX
.
FIGURE 4-3. CURRENT CALIBRATION WINDOW
2. Monitor DVM and click the “+” button to increase and the “–“ to decrease the output current
until the DVM reads as close as possible to maximum rated current (Calculate current as fol-
lows
:
I (Amperes) = DVM reading (Volts) / Shunt Resistance (Ohms).
3. Click the
PRESS TO ACCEPT
button to accept the value. The status display reads
C_0
.
Click the “+” button to increase and the “–“ to decrease the output current until the DVM
reads as close as possible to minimum positive or zero current (Calculate current as follows:
(Amperes) = DVM reading (Volts) / Shunt Resistance (Ohms).
Содержание KLP 10-150-1.2K
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP031308 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 18: ...1 6 KLP 031308 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 19: ...KLP 031308 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 24: ......
Страница 27: ...KLP HV 031308 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......