3-14
KLP031308
after a user-programmable delay, the relay energizes upon a transition
from constant voltage (CV) to constant current (CC) mode.
C->V
: relay energizes immediately upon detection of a system (fatal) error and,
after a user-programmable delay, the relay energizes upon a transition
from constant current (CC) to constant voltage (CV) mode.
RTN
:
Return to
FLT
menu.
MAN -
Relay on/off is controlled manually, directly from the front panel (see below).
Relay state is not retained; relay will be off when unit is turned on
LIST
- Relay on/off is controlled remotely using LIST commands via GPIB or RS 232
interface (see KLP Developer’s Guide).
RTN -
Return to
RELY
menu.
2. To select mode rotate either
VOLTAGE
or
CURRENT
control until status display reads
FLT
(factory),
MAN
(manual), or
LIST
. Then press
DC OUTPUT
switch to save selection.
a. If
FLT
is selected, rotate
VOLTAGE
or
CURRENT
control to select
NORM
,
V->C
, or
C->V
. in the status display.
(1) To select
NORM
, tap the
DC OUTPUT
switch to save selection; unit returns to
RELY
menu.
(2) To select
V->C
or
C->V
, first set the delay. While
V->C
or
C->V
is in the status dis-
play, the
DC AMPERES
display shows the delay in seconds, from 000.5 to 128.0:
NNND
, where
NNN
is an integer from 0 to 128 and
D
will be either .0 or .5 (000.0 is
not permitted). The delay determines the length of time that the unit must be operat-
ing in the new mode before energizing the relay.
The factory default delay time is set to 1.0 seconds to prevent false tripping of the
relay during application of output power to low impedance or highly reactive loads. If
false tripping occurs, this delay may be increased as needed.
To change the delay, press in and rotate either the
VOLTAGE
or
CURRENT
con-
trol; release the control when the
DC AMPERES
display shows the desired delay.
Tap the
DC OUTPUT
switch to save selection; unit returns to
RELY
menu.
NOTE:
Once the relay turns on in
V->C
or
C->V
mode, the relay will go off again when the
unit transitions back to the original mode, or when the output is turned off.
b. To select
MAN
, tap the
DC OUTPUT
switch to save selection. If
MAN
is selected, the
DC
VOLTS
display shows whether the relay is
ON
or
OFF
. To change the state of the relay
press in and rotate either
VOLTAGE
or
CURRENT
control until the desired relay state
is displayed, then release the control. Tap the
DC OUTPUT
switch to immediately set
the relay to the selected state and return to the
RELY
menu.
c. To select
LIST
, tap the
DC OUTPUT
switch to save selection; unit returns to the
RELY
menu
3. Rotate either
VOLTAGE
or
CURRENT
control until status display shows
RTN
and press
DC
OUTPUT
switch to exit
UTIL
menu, then rotate either
VOLTAGE
or
CURRENT
control until
status display shows
EXIT
and press
DC OUTPUT
switch to exit FUNCTION control.
Содержание KLP 10-150-1.2K
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP031308 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 18: ...1 6 KLP 031308 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 19: ...KLP 031308 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 24: ......
Страница 27: ...KLP HV 031308 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......