4-2
KLP 031308
4.3
CALIBRATION USING FRONT PANEL CONTROLS IN LOCAL MODE
For voltage calibration all loads must be disconnected and the sense terminals connected to the
corresponding output terminals. The digital voltmeter will be connected to the output of the
power supply. For current calibration after disconnecting all loads an appropriate shunt resistor
will be connected across the output terminals and the digital voltmeter will be connected across
the sense terminals of the shunt resistor.
1. Turn off power supply and disconnect load from output terminals at the rear of the unit.
2. Verify that power supply is configured for local error sensing (PAR. 2.7.5.1).
3. For voltage calibration connect DVM to (M+, S+) and (M-, S-) of terminal block at the rear of
power supply.
4. For current calibration connect an appropriate shunt resistor (see PAR 4.2) across the output
terminals and connect DVM to shunt resistor terminals.
5. Turn on power supply, exit
SET
mode by tapping either
VOLTAGE
or
CURRENT
control and
using a thin tool (e.g., a paper clip), press the
FUNCTION
switch once until the status display
reads
VIRT
, then rotate either the
VOLTAGE
or
CURRENT
control in either direction until
status display reads
UTIL
.
6. Tap the
DC OUTPUT
switch once, then rotate either the
VOLTAGE
or
CURRENT
control in
either direction until status display reads
CAL
.
7. Tap
DC OUTPUT
switch to enter calibration mode. The status display shows
PASS
. Enter the
password (see PAR. 3.2.5) and tap
DC OUTPUT
switch to accept the number.
8. When the password is accepted, status display reads
V ?
. Tapping either
VOLTAGE
or
CURRENT
control continuously cycles through all the calibration choices. The status display
reads:
V ?
for voltage calibration
C ?
for current calibration,
EXT?
for remote analog cali-
bration, and
EXIT
to save the calibrated results. Tap the
FUNCTION
switch to exit calibration
WITHOUT saving any calibration data (see PAR. 4.3.4). Tap the
DC OUTPUT
switch to
select the displayed choice.
NOTE: Calibrations may be done in any order.
9. Proceed to PAR. 4.3.1 for voltage calibration, PAR 4.3.2 for current calibration, PAR. 4.3.3
for external calibration or PAR. 4.3.4 to exit calibration
4.3.1
VOLTAGE CALIBRATION (LOCAL)
During voltage calibration, the voltage, voltage readback and remote analog voltage readback
are calibrated.
1. Minimum voltage is first; the status display reads
V_0
. Monitor DVM and rotate the
VOLT-
AGE
control to increase or decrease the output voltage until the DVM reads as close as pos-
sible to minimum rated voltage.
2. Tap the
DC OUTPUT
switch to accept the value. The status display then reads
VMAX
. Moni-
tor DVM and rotate the
VOLTAGE
control to increase or decrease the output voltage until the
DVM reads as close as possible to maximum rated voltage.
Содержание KLP 10-150-1.2K
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP031308 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 18: ...1 6 KLP 031308 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 19: ...KLP 031308 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 24: ......
Страница 27: ...KLP HV 031308 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......