KLP031308
3-19
3.3.2
ADDITIONAL FUNCTIONS NOT AVAILABLE FROM THE FRONT PANEL
Some additional functions are described in the KLP Developer’s Guide that are not available
from the front panel:
• Storage of User-programmed Active Settings - 40 memory locations are available to
store and recall active settings of the power supply. The active setting include voltage,
current, protection limits, and output state (ON or OFF) This feature may be also
accessed using the web interface (-1.2K models only, see PAR. 3.3.6.5.3).
• User-determined output Sequences - Up to 100 memory locations are available for pro-
gramming the KLP output using multiple command sequences that can be initiated by a
single LIST command (see KLP Developer’s Guide). This feature may be also accessed
using the web interface (-1.2K models only, see PAR. 3.3.6.5.5).
• External Triggering - If external trigger mode is selected, grounding the EXT_TRG line
(J2 pin 14 to pin 9, 11, 13 or 15) allows SCPI commands to change the output of the
power supply. Refer to the KLP Developer’s Guide for details on the use of triggering.
Kepco provides three instrument drivers, IVI-COM, LabView G and VXI
plug&play
which sim-
plify programming of the KLP power supply via the digital interfaces. These drivers are
described in the KLP Developer’s Guide and can be downloaded from the Kepco web site at:
http://www.kepcopower.com/drivers/
3.3.3
REMOTE PROGRAMMING USING THE IVI-COM INSTRUMENT DRIVER
The IVI-COM driver for KLP is available for download at the Kepco web site www.kepco-
power.com/drivers. Examples are provided in the KLP Developer’s Guide.
3.3.4
REMOTE PROGRAMMING USING THE LABVIEW G INSTRUMENT DRIVER
The LabView G driver for KLP, available for download at the Kepco web site www.kepco-
power.com/drivers, includes a LabView G demonstration program that greatly simplifies pro-
gramming the unit with a Windows™ compatible computer. After running a setup, a “soft” front
panel is visible on the computer monitor that allows all the functionality of the local front panel.
The LabView G soft front panel controls are explained in the KLP Developer’s Guide.
3.3.5
REMOTE PROGRAMMING USING THE VXI
plug&play
INSTRUMENT DRIVER
The VXI
plug&play
driver for KLP, available for download at the Kepco web site www.kepco-
power.com, includes a demonstration program that greatly simplifies programming the unit via
the GPIB or RS 232 ports with a Windows™ compatible computer. After running a setup, a
“soft” front panel is visible on the computer monitor that allows all the functionality of the local
front panel. The VXI
plug&play
soft front panel controls are explained in the KLP Developer’s
Guide.
3.3.6
REMOTE PROGRAMMING USING THE WEB INTERFACE [-1.2K MODELS ONLY]
Remote programming of the KLP via the LAN port can be accomplished using a standard web
browser connected to the internet.
The factory default LAN settings of the KLP are DCHP on, AUTOIP on. When the power supply
is turned on, it tries to find a DCHP server and get an IP address while the LAN indicator blinks
rapidly. If the server is not found after 10 seconds, the unit will use AUTOIP to get an address. If
an IP address is found, the LAN indicator stays on without blinking. If an IP address is not found,
Содержание KLP 10-150-1.2K
Страница 2: ......
Страница 6: ......
Страница 12: ...vi KLP031308 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Страница 18: ...1 6 KLP 031308 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2 R DWG MECH 3010266...
Страница 19: ...KLP 031308 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Страница 24: ......
Страница 27: ...KLP HV 031308 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Страница 40: ......