4-2
BOP-1K 080709
Remote Calibr.
VOLTAGE
EXTERNAL
PROTECTION,
Pos. Voltage
Neg. Voltage
(together)
1. ZERO
2. POSITIVE
External:
1. 0.0V
2. +10.0V
(Analog I/O Port
pin 14+6 ref to
pin12)
N/A
No requirements
1. Automatic internal
measurement
2. Remote Calibration
only.
CURRENT,
INTERNAL
1. ZERO
2. POSITIVE
3. NEGATIVE
Internal
Output Current:
1. 0.0A
2. +Current FS
3. –Current FS
Short with Meas.
Shunt
Automatic CM
1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S)
2. Adjust for zero, posi-
tive and negative FS
CURRENT,
PROTECTION
1. POSITIVE
2. NEGATIVE
Internal
Output Current:
1. +Current FS
2. -Current FS
Short with Meas.
Shunt
Automatic VM
1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S)
2. Adjust for positive
and negative FS.
CURRENT,
EXTERNAL
1. ZERO
2. POSITIVE
External:
1. 0.0V
2. +10.0V
(Analog I/O Port
pin 11 ref to 4)
Output Current:
1. 0.0A
2. +Current FS
Short with Meas.
Shunt
Automatic CM
1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S).
2)Adjust for zero, pos
FS.
CURRENT,
GAIN
1. ZERO
2. POSITIVE
External:
1. 0.0V
2. +10.0V
(Analog I/O Port
pin 11 ref to 4)
Output Current:
1. 0.0A
2. 0.0A
Short with Meas.
Shunt
Automatic CM
1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S)
2. Adjust for zero
(both Zero and Positive)
EXTERNAL
CONTROLS,
PARALLEL
1. ZERO
2. POSITIVE
External:
1. 0.0V
2. +10.0V
(Par/Ser CTRL-IN
Port pin3 ref. to
pin1)
Output Current:
1. 0.0A
2. +Current FS
Short with Meas.
Shunt
Automatic CM
1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S)
2. Adjust for zero, pos.
FS
3. Calibrate unit for
Slave in Parallel
EXTERNAL
CONTROLS,
I_OUT
1. ZERO
2. POSITIVE
INTERNAL
Voltage at Par/Ser
CTRL-IN Port pin 6
ref. to pin1:
1. 0.0V
2. +10.0V
Short with Meas.
Shunt
Automatic CM
1. DVM at Par/Ser
CTRL-IN Port pin 6 ref.
to pin 1
2. Adjust for zero,
+10.0V
3. Calibrate unit for
Master in Parallel
Remote Calibr.
CURRENT
EXTERNAL
PROTECTION,
Pos. Current
Neg. Current
(together)
1. ZERO
2. POSITIVE
External:
1. 0.0V
2. +10.0V
(Analog I/O Port
pins 13, 5 ref to
pin 12)
N/A
No requirements
1. Automatic internal
measurement
2. Remote Calibration.
only.
TABLE 4-1. CALIBRATION SUMMARY (CONTINUED)
Type
Step
Reference Type
and Value
Monitored
Parameter
Output Conditions
Notes
Содержание BOP 1000W
Страница 2: ......
Страница 10: ......
Страница 20: ......
Страница 31: ......
Страница 32: ......
Страница 36: ......
Страница 50: ...FIGURE 2 2 LOAD CONNECTIONS LOCAL SENSING...
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 61: ...FIGURE 2 11 POWER ON DEFAULT SCREEN FOR SLAVE UNIT 2 8 4 OPERATING INSTRUCTIONS FOR MULTIPLE UNIT COMBINATIONS...
Страница 63: ...080709 C 25 FIGURE 2 12 3 X 2 3 SERIES X 2 PARALLEL CONFIGURA TION LOCAL SENSING TYPICAL 2 25 2 26 Blank 24 25...
Страница 64: ......
Страница 68: ...FIGURE 3 3 POWER UP SCREEN SHOWING GRAPHIC METERS VOLTAGE 0000 0...
Страница 125: ......
Страница 128: ......