BOP-1K 080709
4-7
7. Connect a 0V ±0.1mV d-c reference to pin 11 (EXT_REF) referenced to pin 4 (SGND) of the
Analog I/O Port connector (A2A5J6). Set the BOP to zero volts output by sending
CAL:VEXT
ZERO
. Connect the DVM to the BOP output and send
CAL:DATA
commands as needed (see
PAR. 4.3c) to adjust the BOP output until the DVM reads as close to zero as possible within
tolerance specified in Table 4-3 for VOLTAGE ZERO.
NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to
meet published specifications.
8. Replace 0V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6) with a
+10V d-c reference. Set the BOP to maximum positive output voltage by sending
CAL:VEXT
MAX
. Measure the voltage output using the DVM. To adjust, send
CAL:DATA
commands as
needed (see PAR. 4.3b) until the DVM reading is as close as possible above the nominal full
scale value within the tolerance specified in Table 4-3 for +FULL SCALE VOLTAGE.
9. R10V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6)
with a 0V ±0.1mV d-c reference. Set the BOP to zero volts output by sending
CAL:VGA
ZERO
. Connect DVM to the BOP OUT S and OUT COM terminals to measure the output volt-
age. Send
CAL:DATA
commands as needed (see PAR. 4.3c) to adjust the BOP output until
the DVM reads as close to zero as possible within tolerance specified in Table 4-3 for VOLT-
AGE ZERO.
10.Replace 0V reference at pin 11 (EXT_REF) of the Analog I/O Port connector (A2A5J6) with a
+10V d-c reference. Set the BOP to maximum positive output voltage by sending
CAL:VGA
MAX
. Measure the output voltage using the DVM. To adjust, send
CAL:DATA
commands as
needed (see PAR. 4.3b) until the DVM reading is as close to zero as possible within toler-
ance specified in Table 4-3 for VOLTAGE ZERO.
11. Set the BOP to zero volts output by sending
CAL:VLIM ZERO
. Disconnect the +10V d-c ref-
erence voltage from the analog I/O port external reference pin (A2A5J6 pin 11) of the BOP,
then connect 0V ±0.1mV d-c reference voltage from the external voltage source to
VLIM+_EXTERNAL and VLIM–_EXTERNAL (pins 14 and 6) referenced to pin 12 (GND1) of
the BOP analog I/O connector (A2A5J6), then send the
CAL:ZERO
command. This
enables
the BOP to recognize that the voltage at the VLIM_EXTERNAL pins is 0V.
12.C10.0V ±0.1mV d-c reference voltage from the external voltage source to
VLIM+_EXTERNAL and VLIM-_EXTERNAL (pins 14 and 6) referenced to pin 12 (GND1) of
the BOP analog I/O connector (A2A5J6), then send
CAL:VLIM MAX
, then
CAL:ZERO
com-
mands. This enables the BOP to recognize that the voltage at the VLIM+_EXTERNAL pin is
10.0V.
13.Set the BOP to minimum (maximum negative) output by sending
CAL:VLIM MIN
, then send
the
CAL:ZERO
command. This enables the BOP to recognize that the voltage at the VLIM-
_EXTERNAL pin is 10.0V.
14.Connect a 0V ±0.1mV d-c reference to pin 8 (S_IN_SERIAL) of the PAR/SER CONTROL IN
connector (A2A5J3) referenced to COM S terminal at the rear panel (use mating connector
supplied with the unit, P/N 142-0488 to gain access to PAR/SER CONTROL IN connector
pins). Set the BOP to zero volts output by sending
CAL:SER ZERO
. Connect the DVM to the
BOP OUT S and OUT COM terminals and send
CAL:DATA
commands as needed (see PAR.
4.3b) to adjust the BOP output until the DVM reads as close to zero as possible within toler-
ance specified in Table 4-3 for VOLTAGE ZERO.
NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to
meet published specifications.
Содержание BOP 1000W
Страница 2: ......
Страница 10: ......
Страница 20: ......
Страница 31: ......
Страница 32: ......
Страница 36: ......
Страница 50: ...FIGURE 2 2 LOAD CONNECTIONS LOCAL SENSING...
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 61: ...FIGURE 2 11 POWER ON DEFAULT SCREEN FOR SLAVE UNIT 2 8 4 OPERATING INSTRUCTIONS FOR MULTIPLE UNIT COMBINATIONS...
Страница 63: ...080709 C 25 FIGURE 2 12 3 X 2 3 SERIES X 2 PARALLEL CONFIGURA TION LOCAL SENSING TYPICAL 2 25 2 26 Blank 24 25...
Страница 64: ......
Страница 68: ...FIGURE 3 3 POWER UP SCREEN SHOWING GRAPHIC METERS VOLTAGE 0000 0...
Страница 125: ......
Страница 128: ......