4-10
BOP-1K 080709
27.C10.0V ±0.1mV d-c reference voltage from the external voltage source to
ILIM+_EXTERNAL and ILIM-_EXTERNAL, pins 5 and 13, referenced to pin 12 (GND1) of
the BOP analog I/O connector (A2A5J6). Then send
CAL:CLIM MAX
, then
CAL:ZERO
com-
mands. This enables the BOP to recognize that the voltage at the ILIM+_EXTERNAL pin is
10.0V.
28.Set the BOP to minimum (maximum negative) output by sending
CAL:VLIM MIN
, then send
the
CAL:ZERO
command. This enables the BOP to recognize that the voltage at the VLIM-
_EXTERNAL pin is 10.0V.
29.Send
CAL:ZERO
to prepare for calibration of IOUT_M_UNIT (master output) control signal.
Connect the DVM to pin 6 (IOUT_M_UNIT) referenced to pin 1 (SGND) of the PAR/SER
CONTROL IN connector (A2A5J3). Set the BOP to zero output current by sending
CAL:IOUT ZERO
. Send
CAL:DATA
commands as needed (see PAR. 4.3b) to adjust the
BOP output until the DVM reads 0V ±1mV.
30.Set the BOP to maximum positive output current by sending
CAL:IOUT MAX
. With the DVM
still connected to pin 6 (IOUT_M_UNIT) referenced to pin 1 (SGND) of the PAR/SER CON-
TROL IN connector measure the voltage output using the DVM. Send
CAL:DATA
commands
as needed (see PAR. 4.2.1a) until the DVM reading is +10V ±1.0mV.
31.Send
CAL:ZERO
to prepare for calibration of S_IN_PARALLEL (slave input) control signal.
Connect a 0V ±0.1mV d-c reference to pin 3 (S_IN_PARALLEL) referenced to pin 1 (SGND)
of the PAR/SER CONTROL IN connector (A2A5J3). Set the BOP to zero output current by
sending
CAL:PAR ZERO
. Connect the DVM across the sense resistor to measure the BOP
output current and send
CAL:DATA
commands as needed (see PAR. 4.3b) to adjust the
BOP output current until the DVM reads as close to zero as possible within tolerance speci-
fied in Table 4-4 for CURRENT ZERO.
NOTE: Accuracy of the 10V d-c reference must be ±0.1mV in order for the calibrated unit to
meet published specifications.
32.Connect a +10V ±0.1mV d-c reference to pin 3 (S_IN_PARALLEL) referenced to pin 1
(SGND) of the PAR/SER CONTROL IN connector (A2A5J3). Set the BOP to maximum posi-
tive output current by sending
CAL:PAR MAX
. Measure the output current using the DVM.
Send
CAL:DATA
commands as needed (see PAR. 4.3a) until the DVM reading is as speci-
fied in Table 4-3 for +FULL SCALE CURRENT.
33.Save the calibration levels by sending
CAL:SAVE 12/21/2003
. The characters at the end
are optional and represent the calibration date of the power supply. The date entered in this
manner is reported in the
*idn?
query.
4.4
CALIBRATION USING FRONT PANEL KEYPAD IN LOCAL MODE
The following paragraphs describe the screens and keys used in the calibration process.
4.4.1
UNDERSTANDING THE CALIBRATION PROCESS
The following steps describe the calibration screens, and associated keys active during the cali-
bration process. Refer to PAR. 4.4.2 for full calibration procedure.
1. From the power-up screen, press
%
to enter the General Setup menu, then highlight Cali-
bration and press
!
.
Содержание BOP 1000W
Страница 2: ......
Страница 10: ......
Страница 20: ......
Страница 31: ......
Страница 32: ......
Страница 36: ......
Страница 50: ...FIGURE 2 2 LOAD CONNECTIONS LOCAL SENSING...
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 61: ...FIGURE 2 11 POWER ON DEFAULT SCREEN FOR SLAVE UNIT 2 8 4 OPERATING INSTRUCTIONS FOR MULTIPLE UNIT COMBINATIONS...
Страница 63: ...080709 C 25 FIGURE 2 12 3 X 2 3 SERIES X 2 PARALLEL CONFIGURA TION LOCAL SENSING TYPICAL 2 25 2 26 Blank 24 25...
Страница 64: ......
Страница 68: ...FIGURE 3 3 POWER UP SCREEN SHOWING GRAPHIC METERS VOLTAGE 0000 0...
Страница 125: ......
Страница 128: ......