B.56 [SOURce:]LIST:CURRent:APPLy:SWEep? QUERY
LIST:CURR:APPL:SWE?
Syntax:
Short Form: LIST:CURR:APPL:SWE?
Long Form: LIST:CURRent:APPLy:SWEep?
Return Value: <value1>, <value2>(0.01 to 360)
Description:
Returns start
<value1>
and stop
<value2>
angle in degrees for either triangle and sine wave-
forms.
B.57 [SOURce:]LIST:CURRent:POINts? QUERY
LIST:CURR:POIN?
Syntax:
Short Form: LIST:CURR:POIN? Long Form: LIST:CURRent:POINts?
Return Value: <n> where n = number of points
Description:
Identifies the total number of points present in a list and the next location to be filled by
LIST:CURR command.
The LIST:CURR pointer is initially at 0 via LIST:CLE. For each data point
entered by a LIST:CURR command the list pointer is incremented If LIST:CURR:POIN? returns 5, the
LIST:CURR pointer is at 5 indicating there are 6 data points comprising the list. If LIST:VOLT has any
entries, an error message: -221,"Settings conflict" is posted in the error queue.Related Commands:
LIST:CURR. (See example, Figure B-5.)
B.58 [SOURce:]LIST:DWELl COMMAND
LIST:DWEL
Syntax:
Short Form: LIST:DWEL <value> (0.000093 to 0.034),<value>,<value>, . . . to maximum of 5900 val-
ues.
Long Form: LIST:DWELl <value> (0.000093 to 0.034),<value>,<value>, . . . to maximum of 5900 val-
ues.
Description:
Determines how long the main channel parameters will be active.
Sets time value (from 0.000093
to 0.034) in seconds for List:Dwell locations illustrated in Table B-2. These locations are associated
with the corresponding main channel locations illustrated in Table B-2. The main channel is deter-
mined by FUNC:MODE, either current (FUNC:MODE CURR) or voltage (FUNC:MODE VOLT) If
LIST:DWEL is entered for only location 0, that time duration will apply to all steps when either
VOLT:MODE LIST or CURR:MODE LIST is executed. Related Commands: VOLT:MODE,
FUNC:MODE, LIST:CURR, LIST:VOLT, LIST:DWEL?. (See example, Figure B-5.)
B.59 [SOURce:]LIST:DWELl? QUERY
LIST:DWEL?
Syntax:
Short Form: LIST:DWEL?
Long Form: LIST:DWELl?
Return Value: <value>
Description:
Identifies the dwell times entered for the list.
Starting at location established by LIST:QUERy,
returns comma-separated list of up to 16 values indicating the dwell time parameters entered. i.e., the
contents of LIST:DWEL locations of Table B-2. Related Commands: LIST: DWEL, LIST:QUERy. (See
example, Figure B-5.)
B.60 [SOURce:]LIST:DWELl:POINts? QUERY
LIST:DWEL:POIN?
Syntax:
Short Form: LIST:DWEL:POIN?
Long Form: LIST:DWELl:POINts?
Return Value: <value> (0 to 5900)
Description:
Identifies the number of locations for which time values have been entered and the next loca-
tion to be filled by a LIST:DWEL command.
If LIST:DWEL:POIN? returns 6, dwell times have been
entered for locations 0 through 5 and location 6 is the next to be filled by a LIST:DWEL command.
LIST:DWEL, LIST:DWEL:POIN. (See example, Figure B-5.)
B.61 [SOURce:]LIST:QUERy COMMAND
LIST:QUER
Syntax:
Short Form: LIST:QUER <int_value>
Long Form: LIST:QUERy <int_value>
int_value = <n> where n = first location to be queried.
Description:
Determines first location to be queried by LIST:VOLT?, LIST:CURR? or LIST:DWEL? queries.
Related Commands: LIST:VOLT?, LIST:CURR?, LIST:DWEL?. LIST:QUER?. (See example, Figure B-5.)
Содержание BOP 1000W
Страница 2: ......
Страница 10: ......
Страница 20: ......
Страница 31: ......
Страница 32: ......
Страница 36: ......
Страница 50: ...FIGURE 2 2 LOAD CONNECTIONS LOCAL SENSING...
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 61: ...FIGURE 2 11 POWER ON DEFAULT SCREEN FOR SLAVE UNIT 2 8 4 OPERATING INSTRUCTIONS FOR MULTIPLE UNIT COMBINATIONS...
Страница 63: ...080709 C 25 FIGURE 2 12 3 X 2 3 SERIES X 2 PARALLEL CONFIGURA TION LOCAL SENSING TYPICAL 2 25 2 26 Blank 24 25...
Страница 64: ......
Страница 68: ...FIGURE 3 3 POWER UP SCREEN SHOWING GRAPHIC METERS VOLTAGE 0000 0...
Страница 125: ......
Страница 128: ......