BOP HIPWR 080709
1-5
FUNCTION GENERATOR CHARACTERISTICS (Continued)
Types of Basic waveforms
Sine (Sinusoidal waveform)
Triangle (Triangular waveform)
Pos. Ramp (Ramp waveform)
Neg. Ramp (Sawtooth waveform)
Square (50% Duty Cycle Pulse)
Level (DC waveform)
Maximum number of points per basic wave-
form
Local: 3933
Count (Number of repetitions)
Range: 1 to 255
For Count = 0 the waveform is exe-
cuted indefinitely until stop com-
mand is sent
• Local: When the count is complete or if
execution is stopped using
%
, the output
of the unit is set to OFF (Standby indicator
lit). Output must be enabled (press
STANDBY) to execute the waveform
again.
• Remote: When the count is complete or
the waveform is stopped using the
VOLT:MODE FIX command, the output
remains at last output value.
Waveform Name
10 characters maximum
Waveform Parameters
Type: Voltage or Current
Positive Protection Limit:
(Vd-c or Ad-c)
Negative Protection Limit:
(Vd-c or Ad-c)
Count: number
Initial/Repeating
• Type of waveform, voltage or current, can
be selected only for a new waveform and
consequently applies to all waveform seg-
ments. Changing the type for an existing
waveform is not possible, unless the wave-
form is erased first.
• The Limit values (positive/negative) and
the count number apply to all waveform
segments and can be edited at the begin-
ning of a new waveform or later when mod-
ifying a waveform (see PAR. 3.3.9.7).
• All segments except the last may be set to
run once (Initial) or per the Count setting
(Repeating).
Basic waveform parameters
Sine
: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c),
Start Phase (°), Stop Phase (°).
Triangular
: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c), Start
Phase (°), Stop Phase (°).
Ramp (Pos. or Neg.
): Frequency (Hz), Amplitude (Vp-p or Ap-p),
Offset (Vd-c or Ad-c).
Square
: Frequency (Hz), Amplitude (Vp-p or Ap-p), Offset (Vd-c or Ad-c).
Level
: Timing (sec), Amplitude (Vd-c or Ad-c).
• Local: the Start/Stop phases, available for sine and triangle only, can be set or
edited by modifying the existing segment (see PAR. 3.3.9.7).
• Remote: the start/stop phases are global parameters, applicable to all sine/ trian-
gular waveform segments.
• The Start/Stop points are independent and can have any value between 0° and
360°.
TABLE 1-2. BOP GENERAL SPECIFICATIONS (Continued)
SPECIFICATION
RATING/DESCRIPTION
RATING/DESCRIPTION
Содержание BOP 1000W
Страница 2: ......
Страница 10: ......
Страница 20: ......
Страница 31: ......
Страница 32: ......
Страница 36: ......
Страница 50: ...FIGURE 2 2 LOAD CONNECTIONS LOCAL SENSING...
Страница 53: ......
Страница 54: ......
Страница 55: ......
Страница 56: ......
Страница 57: ......
Страница 61: ...FIGURE 2 11 POWER ON DEFAULT SCREEN FOR SLAVE UNIT 2 8 4 OPERATING INSTRUCTIONS FOR MULTIPLE UNIT COMBINATIONS...
Страница 63: ...080709 C 25 FIGURE 2 12 3 X 2 3 SERIES X 2 PARALLEL CONFIGURA TION LOCAL SENSING TYPICAL 2 25 2 26 Blank 24 25...
Страница 64: ......
Страница 68: ...FIGURE 3 3 POWER UP SCREEN SHOWING GRAPHIC METERS VOLTAGE 0000 0...
Страница 125: ......
Страница 128: ......