IEEE-488 Reference
3-100
Attempting to set STATe to ON with TYPE at NONE and MED:STAT at OFF will generate a
“Setting Conflict Error”. Setting STATe to ON during a scan or test sequence will change TCON
to REPeat.
:NTOLerance <n>
[:SENSe[1]]:VOLTage[:DC]:AVERage:ADVanced:NTOLerance <n>
Specify noise tolerance for Volts
[:SENSe[1]]:CURRent[:DC]:AVERage:ADVanced:NTOLerance <n>
Specify noise tolerance for Amps
[:SENSe[1]:RESistance:AVERage:ADVanced:NTOLerance <n>
Specify noise tolerance for Ohms
[:SENSe[1]]:CHARge:AVERage:ADVanced:NTOLerance <n>
Specify noise tolerance for Coulombs
Parameters
<n> = 1 to 100
Specify noise tolerance in percent (+/-)
DEFault
+/-1%
MINimum
0%
MAXimum
+/-100%
Query
:NTOLerance?
Query noise tolerance
:NTOLerance? DEFault
Query *RST default noise tolerance
:NTOLerance? MINimum
Query lowest allowable noise tolerance
:NTOLerance? MAXimum
Query largest allowable noise tolerance
Description
Each measurement function can utilize the advanced filter which adds a noise tolerance window
to the averaging algorithm. These commands are used to set the noise tolerance level of the ad-
vanced filter for the specified function. The :ADVanced[:STATe] commands are used to enable
or disable the advanced filter.
The noise tolerance level is expressed as a percent of the last averaged reading. In general, if the
noise is within this window, the reading is simply based on the normal averaging algorithm. If
a reading is outside this window, then the advanced averaging algorithm is used. See paragraph
2.17.1 for details.
3.19.11 :MEDian Commands
Refer to paragraph 2.17.2 for details on using the median filter.
[:STATe] <b>
[:SENSe[1]]:VOLTage[:DC]:MEDian[:STATe] <b>
Control median; Volts
[:SENSe[1]]:CURRent[:DC]:MEDian[:STATe] <b>
Control median; Amps
[:SENSe[1]]:RESistance:MEDian[:STATe] <b>
Control median; Ohms
[:SENSe[1]]:CHARge:MEDian[:STATe] <b>
Control median; Coulombs
Parameters
<b> =
1 or ON
Enable median filter
0 or OFF
Disable median filter
Query
[STATe]?
Query state of median filter
Description
This command is used to control (on or off) the median filter. When enabled, the instrument will
determine and display the median reading for the specified number of readings. The :RANK
command is used to specify the number of readings used to determine the median.
Note that the median filter can be used with the average filter. When both of these filters are en-
abled, each displayed reading will reflect the median for the average readings.