IEEE-488 Reference
3-118
Effects of positive transitions on the Measurement Event Register:
Positive transition effect on
Measurement event
Measurement Event Register
Reading overflow
Sets B0 when reading exceeds range limits.
Low limit 1
Sets B1 when reading is less than the low limit 1 setting.
High limit 1
Sets B2 when reading is greater than the low limit 1 setting.
Low limit 2
Sets B3 when reading is less than the low limit 2 setting.
High limit 2
Sets B4 when reading is greater than the high limit 2 setting.
Reading available
Sets B5 after a reading is taken and processed.
Reading underflow
Sets B6 when an ohms reading is below the lower range limit.
Buffer available
Sets B7 when the trace buffer has at least two readings.
Buffer half full
Sets B8 when the trace buffer becomes half full.
Buffer full
Sets B9 when the trace buffer becomes full.
Sequence reading available
Sets B10 when an Alternating Polarity test sequence reading
becomes available.
Buffer pretriggered
Sets B11 when the trace buffer pretrigger event occurs.
Out of Limits
Sets B12 when a noise spike occurs.
Fixture lid closed
Sets B13 when fixture lid is closed.
V-source compliance
Sets B14 when V-source goes in compliance.
Figure 3-34
Measurement Transition Filter
RAV
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
(2 )
9
(2 )
8
(2 )
7
(2 )
5
(2 )
4
(2 )
3
(2 )
2
(2 )
1
(2 )
0
32
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Positive Transition
0 = Disable Positive Transition
LL1
2
0/1
Events : VSC = V-Source Compliance
FLC = Fixture Lid Closed
OL = Out of Limits
BPT = Buffer Pretriggered
SRA = Sequence Reading Available
BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
RUF = Reading Underflow
RAV = Reading Available
HL2 = High Limit 2
LL2 = Low Limit 2
HL1 = High Limit 1
LL1 = Low Limit 1
ROF = Reading Overflow
BFL
HL2
HL1
ROF
512
16
4
1
0/1
0/1
0/1
0/1
BHF BAV
LL2
256
128
8
0/1
0/1
0/1
BPT
B11
(2 )
11
2048
0/1
B14
B13
B12
0/1
4096
0/1
B15
(2 )
12
VSC
OL
RUF
16384
(2 )
14
(2 )
6
256
0/1
A) Positive Transition (PTR) Register
RAV
B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
(2 )
9
(2 )
8
(2 )
7
(2 )
5
(2 )
4
(2 )
3
(2 )
2
(2 )
1
(2 )
0
32
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Negative Transition
0 = Disable Negative Transition
LL1
2
0/1
Events : VSC = V-Source Compliance
FLC = Fixture Lid Closed
OL = Out of Limits
BPT = Buffer Pretriggered
SRA = Sequence Reading Available
BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
RUF = Reading Underflow
RAV = Reading Available
HL2 = High Limit 2
LL2 = Low Limit 2
HL1 = High Limit 1
LL1 = Low Limit 1
ROF = Reading Overflow
BFL
HL2
HL1
ROF
512
16
4
1
0/1
0/1
0/1
0/1
BHF BAV
LL2
256
128
8
0/1
0/1
0/1
BPT
B11
(2 )
11
2048
0/1
B14
B13
B12
0/1
4096
0/1
B15
(2 )
12
VSC
OL
RUF
16384
(2 )
14
(2 )
6
256
0/1
A) Negative Transition (NTR) Register
SRA
(2 )
10
1024
0/1
SRA
(2 )
10
1024
0/1
FLC
8192
(2 )
13
0/1
FLC
8192
(2 )
13
0/1