Front Panel Operation
2-70
2.14.3 Running the selected test
Perform the following steps to run the selected test:
1. Enable zero check and make sure the V-Source is in
standby (OPERATE LED off).
2. Connect and configure the Model 6517A for the desired
test as explained in paragraph 2.14.1.
3. Select and configure the desired test as explained in
paragraph 2.14.2.
4. Press the SEQ key. The selected test will be displayed.
5. Press ENTER to arm the test. When the selected trigger
source event occurs, zero check will disable and the test
will run.
Table 2-20 (cont.)
CONFIGURE SEQUENCE menu structure
Menu item
Description
Default
parameter
ALT-POLARITY
V-OFS
V-ALT
MEAS-TIME
DISCARD RDGS
STORE RDGS
SIR
SUR-INSUL-RES-TEST
BIAS V
BIAS-TIME
MEAS-V
MEAS-TIME
Alternate Polarity Test:
Specify offset voltage.
Specify alternating voltage.
Specify measurement time.
Specify discarded readings.
Specify readings to store.
Surface Insulation Resistance Test:
Specify bias voltage.
Specify bias time.
Specify measurement voltage
Specify measurement time.
0V
10V
15 sec
3
1
+50V
1 sec
+100V
1sec
SWEEP
SQUARE-WAVE
HI-LEVEL
HI-TIME
LO-LEVEL
LO-TIME
CYCLE COUNT
STAIRCASE
START
STOP
STEP
STEP TIME
Sweep Tests:
Square Wave Sweep Test:
Specify high level voltage.
Specify time at high level.
Specify low level voltage.
Specify time at low level.
Specify number of cycles.
Staircase Sweep Test:
Specify start voltage.
Specify stop voltage.
Specify step voltage.
Specify step time.
+1V
1sec
-1V
1sec
10
+1V
+10V
+1V
1sec
CONTROL
MANUAL
IMMEDIATE
LID-CLOSURE
GPIB
EXTERNAL
TRIGLINK
Select trigger source to start test:
Start when TRIG key pressed.
Start immediately.
Start when test fixture lid closed.
Start on GPIB trigger (GET or *TRG).
Start when external trigger received.
Start when trigger is received via the speci-
fied Trigger Link line.
Manual
Line #1
6. When the test is finished, zero check will stay disabled
and the V-Source will go into standby.
7. The measured readings for the test are stored in the buff-
er. To access these readings, press RECALL.
Notes:
1. If the IMMEDIATE trigger source is selected, the test
will start immediately after it is armed. With any other
trigger source (except LID CLOSURE) selected, the test
can be started by pressing TRIG.
2. While a test is armed or running, the flashing “SEQ”
message is displayed on the Model 6517A.