IEEE-488 Reference
3-149
:TSOurce <name>
:TSEQuence:TSOurce <name>
Select control source to start test
Parameters
<name> = MANual
Manual control source
IMMediate
Immediate control source
BUS
Bus trigger control source
EXTernal
External trigger control source
TLINk
Trigger link control source
LCLosure
Test fixture lid closure
Query
:TSOurce?
Query control source to start test
Description
This command is used to select the event that will start the selected test sequence. When the se-
lected control source event occurs, the test will start.
With MANual selected, the test will start when the SEQ key is pressed. With IMMediate select-
ed, the test will start as soon as the instrument is taken out of the idle state. The :INITiate and
:INITiate:CONTinuous ON commands will take the instrument out of the idle state. Note that
if the instrument is already in the continuous measurement mode (not in idle), the test will start
as soon as IMMediate is selected.
With BUS selected, the test will start when a bus trigger (*TRG or GET) is sent over the bus.
With EXTernal selected, the test will start when an external trigger is received via the EXT
TRIG IN connector on the rear panel.
With TLINk selected, the test will start when an external trigger is received via the TRIG LINK
connector on the rear panel. The :TLINe command is used to select which trigger link line to
use for the input trigger.
With LCLOsure selected, the test will start when the test fixture lid is closed. This will only
work for test fixtures (such as the Models 8002A and 8009) that incorporate the safety interlock
feature. If any sequence is armed when this command is sent, it will not take effect until the next
sequence arm event.
:TLIne <NRf>
:TSEQuence:TLIne <NRf>
Specify trigger link line
Parameters
<NRf> = 1 Line #1
<NRf> = 4 Line #4
2 Line #2
5 Line #5
3 Line #3
6 Line #6
Query
:TLINe?
Query trigger link line
Description
This command is used to specify the input trigger link line for the TLINk control source. This
command is not used if a different control source is selected (see :TSOUrce).
3.25.2
:STARt <NRf>
:TSEQuence:DLEakage:STARt <NRf>
Start voltage; diode leakage test
:TSEQuence:STSWeep:STARt <NRf>
Start voltage; staircase sweep test
Parameters
<NRf> = -100.0 to 100.0
Volts; 100V range
-1000 to 1000
Volts; 1000V range