IEEE-488 Reference
3-108
:LIMit:STATe <b>
:SOURce:VOLTage:LIMit:STATe <b> Control voltage limit
Parameters
<b> =
0 or OFF
Disable voltage limit
1 or ON
Enable voltage limit
Query
:STATe?
Query state of voltage limit
Description
This command is used to enable or disable voltage limit. When enabled, the programmed volt-
age limit is in effect. When disabled, voltage limit is determined by the selected range.
Attempts to enable voltage limit when the present V-Source level (amplitude) exceeds the limit
value will generate a settings conflict error.
:MCONnect <b>
:SOURce:VOLTage:MCONnect <b> Control LO connection
Parameters
<b> =
ON or 1
Enable LO connection
OFF or 0
Disable LO connection
Query
:MCONnect?
Query LO connection
Description
This command is used to control the internal V-Source LO to Ammeter LO connection. When
enabled, V-Source LO is connected to Ammeter LO. This connection simplifies the connection
requirements for ohms measurements and source voltage measure current measurements. When
disabled, this connection is broken.
:CURRent Commands
:RLIMit:STATe <b>
:SOURce:CURRent:RLIMit:STATe <b>
Control resistance current limit
Parameters
<b> =
0 or OFF
Disable resistance current limit
1 or ON
Enable resistance current limit
Query
:STATe?
Query state of resistance current limit
Description
This command is used to enable or disable resistance current limit for the V-Source. When en-
abled, a 1M
Ω
resistor is placed in series with the HI V-SOURCE OUT lead to extend the current
limit capabilities of the V-Source. When disabled, normal current limit occurs at 1mA.
:LIMit[:STATe]?
:SOURce:CURRent:LIMit[:STATe]?
Query current compliance.
This command is used to determine if the V-Source has reached its current compliance limit of
1mA. The response message is explained as follows:
0 = Current compliance not reached
1 = Current compliance reached