![background image](http://html2.mh-extra.com/html/keithley/6517a/6517a_user-manual_4066185258.webp)
IEEE-488 Reference
3-107
Note that the amplitude level is limited by the selected range and programmed voltage limit (see
:RANGe and :LIMit). When on the 100V range, attempts to set the amplitude to a level that ex-
ceeds the 100V range will result in a Setting Conflicts error. Attempts to set an amplitude that
exceeds a programmed voltage limit will set the amplitude to that limit.
The MINimum and MAXimum parameters are also affected by range and voltage limit. Setting
the V-Source to the MINimum (or MAXimum) amplitude selects the greatest negative (or pos-
itive) value allowed by the programmed limit or range.
The V-Source output (operate and standby) is controlled from the :OUTPut1 subsystem (see
paragraph 3.17).
:RANGe <n>
:SOURce:VOLTage:RANGe <n>
Select V-Source range
Parameters
<n> =
0 to 1000
Specify output in volts
DEFault
100V range
MINimum
100V range
MAXimum
1000V range
Query
:RANGe?
Query voltage range
:RANGe? DEFault
Query *RST default range
:RANGe? MINimum
Query minimum range
:RANGe? MAXimum
Query maximum range
Description
This command is used to select the optimum range for the V-Source. Specifying the output volt-
age value automatically selects the most sensitive range. For example, specifying a value
≤
100
selects the 100V range. Specifying a value >100 selects the 1000V range.
You cannot select the 100V range if the programmed amplitude value is greater than 100. A set-
ting conflict error will result.
:LIMit[:AMPLitude] <n>
:SOURce:VOLTage:LIMit[:AMPLitude] <n>
Set voltage limit
Parameters
<n> =
0 to 1000
Set voltage limit
DEFault
1000V
MINimum
0V
MAXimum
1000V
Query
:LIMit?
Query voltage limit
:LIMit? DEFault
Query *RST default limit
:LIMit? MINimum
Query minimum limit
:LIMit? MAXimum
Query maximum limit
Description
This command is used to set an absolute value limit for the V-Source. For example, setting a
limit of 75V limits the V-Source output to ±75V.
The specified voltage limit is only in effect if the limit is enabled (see :LIMit:STATe). If dis-
abled, the limit is simply dictated by the selected range.