IEEE-488 Reference
3-92
Parameters
<n> =
0 to 210
Range limit for volts
0 to 21e-3
Range limit for amps
DEFault 210
(volts)
21e-3 (amps)
MINimum
0 (volts and amps)
MAXimum
Same as DEFault
Query
:ULIMit?
Query upper range limit :ULIMit?
DEFault
Query *RST default upper limit :ULIMit?
MINimum
Query lowest allowable upper limit :ULIMit?
MAXimum
Query largest allowable upper limit
:LLIMit?
Query lower range limit :LLIMit?
DEFault
Query *RST default lower limit :LLIMit?
MINimum
Query lowest allowable lower limit :LLIMit?
MAXimum
Query largest allowable lower limit
Description
Automatic range changes slow down the reading rate. By setting upper and lower measurement
range limits, you can prevent inadvertent range changes to ranges that you are not going to use.
The upper range limit is selected by specifying the maximum expected reading that you expect
to measure. For example, if the maximum expected reading is approximately 1mA, let the pa-
rameter (<n>) = 0.001 (or 1e-3). The lowest (most sensitive) range that can measure 1mA is the
2mA range. Thus, the 2mA range is selected as the upper range limit. With this upper limit, the
instrument cannot uprange to the 20mA measurement range.
The lower range limit is selected by specifying the lowest expected reading that you expect to
measure.
Note: See next command to set the autorange limit for coulombs.
:LGRoup <name>
[:SENSe[1]]:CHARge:RANGe:AUTO:LGRoup <name>
Select coulombs autorange limit
Parameters
<name> = HIGH
200nC to 2µC autorange limit
LOW
2nC to 20nC autorange limit
Query
:LGRoup?
Query coulombs autorange limit
Description
This command is used to set the autorange limit for the coulombs function. With HIGH selected,
the instrument will only autorange between the two high coulombs ranges (200nC and 2µC).
With LOW selected, the instrument will only autorange between the two low coulomb ranges
(2nC and 20nC).
Ohms Ranges:
[:UPPer] <n>
[:SENSe[1]:RESistance[:AUTO]:RANGe[:UPPer] <n>
Set measurement range for Auto V-Source Ohms
[:SENSe[1]:RESistance:MANual:CRANge[:UPPer] <n>
Set measurement range for Manual V-Source Ohms