IEEE-488 Reference
3-54
:STATus
:QUEStionable
[:EVENt]?
:ENABle <NRf>
:ENABle?
:PTRansition <NRf>
:PTRansition?
:NTRansition <NRf>
:NTRansition?
:CONDition?
:PRESet
:QUEue
[:NEXT]?
:ENABle <list>
:ENABle?
:DISable <list>
:DISable?
:CLEar
Path to control questionable status registers:
Read the event register.
Program the enable register.
Read the enable register.
Program the positive transition register.
Read the positive transition register.
Program the negative transition register.
Read the negative transition register.
Read the condition register.
Return status registers to default states.
Path to access error queue:
Read the most recent error message.
Specify error and status messages for queue.
Read the enabled messages.
Specify messages not to be placed in queue.
Read the disabled messages.
Clears all messages from Error Queue.
(Note 2)
(Note 3)
(Note 4)
(Note 5)
(Note 6)
(Note 7)
(Note 7)
√
√
√
√
√
√
√
√
√
√
√
√
√
√
3.21.1
3.21.2
3.21.3
3.21.4
3.21.5
3.21.6
3.21.7
Notes:
1. Commands in this subsystem are not affected by *RST and :SYSTem:PRESet. The effects of cycling power, *CLS and :STATus:PRESet are explained
by the following notes.
2. Event Registers:
Power-up and *CLS – Clears all bits of the registers
:STATus:PRESet – No effect.
3. Enable Registers: Power-up and :STATus:PRESet – Clears all bits of the registers
*CLS – No effect
4. PTR Registers:
Power-up and :STATus:PRESet sets all bits of the registers
*CLS – No effect
5. NTR Registers:
Power-up and :STATus:PRESet clears all bits of the registers
*CLS – No effect
6. Error Queue:
Power-up and *CLS – Clears the Error Queue
:STATus:PRESet – No effect
7. Enable/Disable Error Queue Messages: Power-up – Clears list of messages
*CLS and :STATus:PRESet – No effect
Table 3-12 (Continued)
STATus command summary
Command
Description
Default
parameter
SCPI
Ref.