Front Panel Operation
2-63
figured from the CONFigure SEQUENCE menu (DEV-
CHAR; RESISTOR). See paragraph 2.14.2 for details.
6517A
HI
6517A
V-Source
A
+
-
6517A
Ammeter
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
7078-TRX Cable
INPUT
250V PEAK
COMMON
TRIGGER
LINK
IN
OUT
!
LINE RATING
90-134VAC
180-250VAC
50, 60, 400HZ
55VA MAX
LINE FU
SLOWBL
1/2A, 25
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
PREAMP OUT
250V PEAK
V SOURCE
LO
HI
Note: Ammeter LO internally
connected to V-Source
LO (See Paragraph 2.9.1).
A) Connections
HI
LO
HI
LO
B) Equivalent Circuit
Cable
Shield
Insulator
Center
Conductor
Cable
Resistance
Figure 2-50
Connections; cable insulation resistance test
(SOURCE V2) is then applied and, after the next delay (DE-
LAY 2), a second resistance measurement is made. The
Model 6517A then automatically calculates the voltage coef-
ficient and stores it in the buffer. This test is selected and con-